If you set Fluke DAQ Software for general network operation, you must set the
network type of each device to general. You will need to enter an IP address, and
possibly a socket port, subnet mask and gateway address into each device. Get this
information from your network administrator.
Table 4. Reviewing and Setting the Network Type to General

Press the COMM key to review the network type, or press and hold the COMM
key for 3 seconds to set the network type.

Press the up/down arrow keys until nEt (Network) appears in the primary

display (comm appears in the secondary display).

Press the ENTER key. nEt appears in the secondary display and ISo (isolated
network) or gEn (general network) is in the primary display.

To set the network type to general, press the up/down arrow keys to

select gEn (current setting appears bright).

Press the ENTER key. (Pressing any other function key will cancel set
operations.) This displays the current Socket Port.

Press the ENTER key. This displays the first digit of the Internet Protocol
address (segment IP:0).

Press the ENTER key to exit. You must set an IP address and Socket Port
when using a general network.
Getting Started
Configuring the 268XA
9