Press "SET" to set the baud rate to the new value or "EXIT" to abort the opera-
tion and skip to the next parameter in the menu.
7.8.1

Sample Period

The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure-
ments transmitted from the serial interface. If the sample rate is set to 5, the in-
strument transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. The sample period is indicated by,
SPEr
Press "SET" to choose to set the sample period. The current sample period
value will be displayed.
SP= 1
Adjust the value with "UP" or "DOWN" and then use "SET" to set the sample
rate to the displayed value. Press "EXIT" to continue without changes.
SP= 60
7.8.1.1

Duplex Mode

The next parameter is the duplex mode. The duplex mode may be set to full du-
plex or half duplex. With full duplex any commands received by the calibrator
via the serial interface are immediately echoed or transmitted back to the device
of origin. With half duplex the commands are executed but not echoed. The du-
plex mode parameter is indicated by,
dUPL
Press "SET" to access the mode setting.
d=FULL
The mode may be changed using "UP" or "DOWN" and pressing "SET".
d=HALF
New baud rate
Serial sample period parameter
Current sample period (seconds)
New sample period
Serial duplex mode parameter
Current duplex mode setting
New duplex mode setting