Fluke 9140 User Manual page 25

Dry-well calibrator
Hide thumbs Also See for 9140:
Table of Contents

Advertisement

6 General Operation
Calibrating Probes
Using the same hole for the reference thermometer and the test probe may have
better results. This however requires switching the probes which takes more
time. One must allow a few minutes after inserting the probes for the tempera-
ture to stabilize before making measurements. Because of temperature varia-
tions along the length of the well, best results are obtained when comparing
probes of similar construction and inserting them the same depth into the well.
19