Fluke 9140 User Manual page 41

Dry-well calibrator
Hide thumbs Also See for 9140:
Table of Contents

Advertisement

9140 Dry-well Calibrator
User's Guide
Table 2 9140 controller communications commands
Command Description
Display Temperature
Read current set-point
Set current set-point to n
Read temperature
Read temperature units
Set temperature units:
Set temperature units to Celsius
Set temperature units to
Fahrenheit
Read scan mode
Set scan mode
Read scan rate
Set scan rate
Read hold
Secondary Menu
Read proportional band setting
Set proportional band to n
Read heater power
(duty cycle)
Configuration Menu
Calibration Menu
Read R0 calibration parameter
Set R0 calibration parameter to n r[0]=n
Read ALPHA calibration
parameter
Set ALPHA calibration parameter
to n
Read DELTA calibration
parameter
Read DELTA calibration
parameter
Operating Parameters Menu
Serial Interface Menu
Read serial sample setting
Set serial sampling setting to n
seconds
Set serial duplex mode:
Set serial duplex mode to full
Set serial duplex mode to half
36
Command
Command
Format
Example
s[etpoint]
s
s[etpoint]=n
s=350
t[emperature]
t
u[nits]
u
u[nits]=c/f
u[nits]=c
u=c
u[nits]=f
u=f
sc[an]
sc
sc[an]=on/off
sc=on
sr[ate]
sr
sr[ate]=n
sr=1.1
ho[ld]
ho
pr[op-band]
pr
pr[op-band]=n
pr=8.83
po[wer]
po
r[0]
r
r=100.7
a[lpha]
al
a[lpha]=n
al=0.003865
d[elta]
de
d[elta]=n
de=1.37
sa[mple]
sa
sa[mple]=n
sa=0
du[plex]=f[ull]/h[alf]
du[plex]=f[ull]
du=f
du[plex]=h[alf]
du=h
Returned
Returned
Example
set: 9999.99 {C or F}
set: 150.00 C
t: 9999.9 {C or F}
t: 55.6 C
u: x
u: C
sc: {ON or OFF}
sc: ON
srat: 99.9 C/min
srat:12.4 C/min
ho: open/closed, 99.9 {C
ho: open,
or F}
30.5 C
pb: 999.9
pb: 15.9
po: 999.9
po: 6.5
r0: 999.999
r0: 100.7
al: 9.999999
al: 0.003865
de: 1.50
de: 9.9999
sa: 9
sa: 1
Acceptable
Values
Instrument
Range
C or F
ON or OFF
.1 to 99.9
Depends on
Configuration
97.0 to 105
.002 to .006
0–3.0000
0 to 999
FULL or HALF

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents