Troubleshooting; Troubleshooting Problems, Possible Causes, And Solutions - Fluke 9140 User Manual

Dry-well calibrator
Hide thumbs Also See for 9140:
Table of Contents

Advertisement

Solutions
12
12.1

Troubleshooting

If problems arise while operating the 9140, this section provides some sugges-
tions that may help you solve the problem. A wiring diagram is also included.
Troubleshooting Problems, Possible Causes,
and Solutions
In the event that the instrument appears to function abnormally, this section
may help to find and solve the problem. Several possible problem conditions
are described along with likely causes and solutions. If a problem arises, please
read this section carefully and attempt to understand and solve the problem. If
the problem cannot otherwise be solved, contact a Hart Scientific Authorized
Service Center (see Section 1.3) for assistance. Be sure to have the model
number, serial number, voltage, and problem description available.
Problem
Possible Causes and Solutions
Incorrect tempera-
Incorrect R0, ALPHA, and DELTA parameters. Find the value for R0, ALPHA, and
ture reading
DELTA on the Report of Calibration. Reprogram the parameters into the instrument
(see Section , Calibration Parameters). Allow the instrument to stabilize and verify
the accuracy of the temperature reading.
Controller locked up. The controller may have locked up due to a power surge or
other aberration. Initialize the system by performing the Factory Reset Sequence.
Factory Reset Sequence. Hold the SET and EXIT buttons down at the same time
while powering up the instrument. The instrument displays shows '-init-', the model
number, and the firmware version. Each of the controller parameters and calibration
constants must be reprogrammed. The values can be found on the Report of
Calibration.
The instrument
Incorrect scan and scan rate settings. The scan and scan rate settings may be
heats or cools too
set to unwanted values. Check the Scan and Scan Rate settings. The scan may be
quickly or too
off (if the unit seems to be responding too quickly). The scan may be on with the
slowly
Scan Rate set low (if unit seems to be responding too slowly).
Improper line voltage. Verify that the voltage reading in the bottom of the unit
matches the source voltage.
Unstable display
Wait. Allow the instrument to stabilize for a few minutes.
Proportional band may be incorrect. Refer to the proportional band on the Report
of Calibration.
Troubleshooting Problems, Possible Causes, and
12 Troubleshooting
49