9-3. Restoring to the factory setting values

If the gravity acceleration value or calibration weight value is changed unintentionally,
restore those values to the factory setting values, as follows.
Make sure that the scale is in the weighing
mode (displaying "kg", "lb", "oz" or "pcs").
1. Press and hold the TARE switch until
the gravity acceleration value appears,
and release the switch.
2. Press the
ClrC
display "
.
3. Press and hold the UNITS switch until
ClrCno
"
" appears blinking.
4. Press the PRINT switch.
ClrCno
"
" changes to "
ClrCgo
5. When "
TARE
switch. The factory setting
values
are
displayed.
To finish the restoring procedure, turn the
power OFF.
UNITS
switch twice and
" .
ClrCgo
".
" is displayed, press the
restored
and
"
(more than 4 seconds)
end
"
is
19