Compares the weight value with the upper and lower limit values and shows the results
of the comparison by
The formula to compare is as follows:
< Lower limit value ≤
1LO
Select comparison conditions that suit your own application. (Refer to "Comparator mode: Cp"
in "10-2. Function list".)
0: No comparison (comparator function disabled).
1: To compare all data.
2: To compare all stable data.
3: To compare all data which are more than +4d or less than -4d.
4: To compare stable data which are more than +4d or less than -4d.
5: To compare all data which are more than +4d.
6: To compare stable data which are more than +4d.
d = minimum display (Refer to "13-1. Specifications".)
Also in the counting mode, "d" is equal to the minimum display of kg mode.
The upper and lower limit values are maintained even if the power is turned OFF.
The upper and lower limit values are common to the weighing and counting mode.
Ignore the decimal point when setting the upper limit or lower limit value.
For example, when the weighing capacity is 30 kg and the setting value is "001000":
Display mode
Normal resolution kg
High resolution kg
Normal resolution lb
High resolution lb
Normal resolution oz
High resolution oz
Counting mode

8. COMPARATOR

on the display.
HI
OK
LO
≤ Upper limit value <
OK
Limit value
10.00 kg
1.000 kg
10.00 lb
10.00 lb
100.0 oz
100.0 oz
1000 pcs
15
HI
Capacity / Minimum display
30.00 kg / 0.01 kg
30.000 kg / 0.005 kg
66.00 lb / 0.02 lb
66.00 lb / 0.01 lb
1050.0 oz / 0.5 oz
1050.0 oz / 0.2 oz