Table D-1. Computer Control Commands (cont.)
Description
Perform Leak Test at
XXXmmHg
Perform Pressure Source at
XXXmmHg
Performs Relief Valve Test
Pressure Gauge
D-4
Command
[LEAK,XXX]
50<=XXX<=400
[PSOURCE,XXX]
50<=XXX<=400
[BLEED_SYSTEM]
[RELIEF,XXX]
50<=XXX<=400
[GAUGE]
Returned
String
Returns ACK
then Leak Rate.
Returns Leak
Rate after 60
seconds,
followed by CR-
LF, or NAK if
XXX out of
range•
ACK or NAK if
XXX out of
range
ACK
Returns ACK
then Peak
Pressure,
followed by CR-
LF, or NAK if
XXX out of
range
Returns ACK
then pressure
port value,
followed by CR-
LF