BP Pump 2
Operators Manual
All tests begin from the Main menu:
Fluke Biomedical BP Bump 2
PERFORM
SIMULATION
After pressing the PRESSURE TESTS soft key to display the Pressure Tests
menu, you can access the pressure tests described below:
PRESSURE
LEAK TEST
Pressure Leak Test
The Pressure Leak Test pressurizes a pneumatic system to an operator-defined
target pressure (labeled Setpoint) up to 400 mmHg (53.3 kPa) and then measures
the loss of pressure over time.
To assess leakage, take the following actions:
1. Define the Setpoint by using the soft keys to follow the menu path shown
below.
PRESSURE TESTS C PRESSURE LEAK TEST C SETUP
The Leak Test (Setpoint) screen appears.
Leak Test
2. Adjust the Setpoint by entering numbers with the number keys. The cursor
moves to the next position after a number is entered
3-8
PRESSURE
AUTO
TESTS
SEQUENCE
PRESSURE
STATIC
RELIEF
PRESSURE
Setpoint: 250
SETUP
fas31.eps
fas35.eps
fas41.eps