shape of the pulses are controlled by the microprocessor driving the step motor.
The home position of the piston is detected by an optical interrupter. If the
optional high-accuracy pressure sensor is installed, it also measures the pressure in
the main manifold and is also controlled by the microprocessor.

Configurations for Devices Under Test (DUT)

Connect the Tester to the NIBPM unit with the tubing and fittings (part number
2391882) supplied with the Tester in the desired configurations, as shown in
Figures 3-2 through 3-6.
Basically, configurations involve connecting the NIBPM unit to the Tester directly
or connecting the unit, a cuff, and the Tester, using a "T" connector. When
connected properly, the tubing forms a closed system with the components.
Single Hose: Internal Cuff
NIBP Monitor
120
93
80
Figure 3-2.Connecting Tester to Single-hose NIBP Monitor
3-4
80
(Int Cuff)
BP Pump 2
NON-INVASIVE BLOOD PRESSURE MONITOR ANALYZER
TESTS AND SIMULATIONS
PRESSURE
PRESSURE
STATIC
LEAK
RELIEF
PRESSURE
PATIENT
STANDARD BP
CONDITIONS
ARRHYTHMIAS
RESPIRITORY
NEONATE
ARTIFACT
WRIST
USER DEFINED
ENT
fas16.eps