Fluke BP Pump 2 Operator's Manual page 46

Nibp simulator and tester
Table of Contents

Advertisement

BP Pump 2
Operators Manual
Put the NIBP monitor in "Calibrate" or "Service" mode to close the vent
valve, allowing the Tester to inflate the pneumatic system. Refer to the
NIBP monitor's service manual to find the method for entering "Service"
mode.
To assess the effectiveness of the relief valve, do the following:
1. Use the soft keys to follow the menu path shown below.
PRESSURE TESTS C PRESSURE RELIEF C SETUP
The Relief Valve Test (Setpoint) screen appears.
Relief Valve Test
2. If necessary, adjust the Setpoint, which defaults to 380 mmHg, by entering
numbers with the number keys. The cursor moves to the next position after a
number is entered.
3. Press ENT to display the Relief Valve Test screen.
Relief Valve Test
4. Press the VENT soft key to release any unwanted pressure in the system
before performing the test. This feature vents the system for approximately
five seconds and can be repeated as needed to return pressure to zero.
5. Press the <START> soft key to make the Tester deliver air to the system.
While the Tester is delivering air to the system, the current pressure
(Measured) and peak pressure are being monitored.
If the Setpoint is reached and the monitor does not release the pressure, the
message No Relief Detected appears on the display.
3-10
Note
Setpoint: 380
Peak : _._ _
Setpoint : 380
Measured : 0
SETUP
VENT
mm Hg
mm Hg
mm Hg
mm Hg
<START>
fas02.eps
fas03.eps

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents