Maintenance and Support

The Tester requires little maintenance or special care; however, it is a
calibrated measuring instrument and should be treated as such. The following
describes how to maintain the Tester and the method of company support.

Avoiding Damage

Do not drop the instrument or subject it to any mechanical abuse that could
cause a shift in the calibrated settings.
To avoid damage to the Tester or adverse affects on its
performance, follow these guidelines:
Do not expose the system to temperature extremes.
Ambient temperatures should remain between 15 and
40 °C. System performance may be adversely affected
if temperatures fluctuate above or below this range.
Do not apply pressures greater than 400 mmHg (53
kPa) to the pressure port.

Cleaning

Clean the exterior of the Tester occasionally with a cloth dampened with a
mild detergent solution. Take care to keep liquids out of the pressure port.
To avoid damage to the Tester or adverse affects on its
performance, clean it only by gently wiping down with a
clean, lint-free cloth dampened with a mild detergent
solution. Do not immerse the unit.
Carefully wipe down the hose adapters and inspect them for damage and
deterioration of the tubing and fittings.
2-10