Arrhythmias; Patient Condition Simulations - Fluke BP Pump 2 Operator's Manual

Nibp simulator and tester
Table of Contents

Advertisement

BP Pump 2
Operators Manual
Patient Condition
Healthy
Cuff: External
OPTIONS
2. Press the OPTIONS soft key to scroll through the simulation choices.
Parameters for Patient Condition simulations are shown in Table 3-3.
Patient Condition
Healthy Heart
Weak Pulse
Mild Exercise #1
Strenuous Exercise #2
Obese Subject
Geriatric Subject
Tachycardia
Bradycardia
3. Press the CUFF soft key to select Internal Adult or External cuff.
4. Press Start on the NIBP unit.
5. Compare values on the NIBP unit with those on the Tester.
6. Press HOME to return to the Main menu.

Arrhythmias

These waveforms cause erratic readings on some NIBPMs. The blood pressure
determination strongly depends on exactly what is happening with the subject's
blood pressure when the cuff pressure is at a particular level. Some NIBPMs
pause until they detect two or more equivalent beats. The pattern of step deflations
and the measured blood pressure depend on which beats occur during each step of
the cuff pressure.
3-16
CUFF
Table 3-3. Patient Condition Simulations
Blood Pressure
(mmHg) (MAP)
120/80 (93)
110/80 (90)
140/90 (106)
140/90 (106)
120/80 (93)
150/110 (123)
120/105 (110)
120/60 (80)
ready ....
120/
80 (93)
75 BPM
0.68cc
Heart Rate
Pulse volume
(bpm)
75
95
120
162
90
95
130
45
fas08.eps
(cc)
0.68
0.50
1.00
1.40
0.50
0.40
0.40
1.10