Page 2
Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility.
Fluke Biomedical. Copyright Release Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Page 4
United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.
Page 5
Fluke Biomedical. Changes made to the information in this document will be incorpo- rated in new editions of the publication. No responsibility is assumed by Fluke Biomedi- cal for the use or reliability of software or equipment that is not supplied by Fluke Bio- medical, or by its affiliated dealers.
Table of Contents Chapter Title Page Introduction and Specifications..........1-1 Introduction ..................1-3 Key Features ..................1-3 General Safety Considerations.............. 1-4 Symbols .................... 1-4 Warnings and Cautions ..............1-5 Instrument Familiarity ................1-7 Powering Up the Tester ................ 1-10 Specifications..................1-10 Accessories ...................
Page 8
BP Pump 2 Operators Manual Pressure Tests..................3-8 Pressure Leak Test................3-8 Pressure Relief Test................3-10 Pressure Source Test ................ 3-11 Pressure Gauge Test ................. 3-12 Simulations................... 3-13 Standard BP..................3-14 Patient Conditions ................3-15 Arrhythmias..................3-16 Respiratory Artifacts ................ 3-18 Neonate ....................
Page 9
List of Tables Table Title Page 1-1. Symbols ....................1-4 1-2. Top and Side Panel Components ............1-8 1-3. Number Key Functions................. 1-9 1-4. Standard Accessories ................1-12 1-5. Optional Accessories ................1-12 3-1. Conversion Factors ................3-7 3-2. Standard Blood Pressure Simulations ........... 3-15 3-3.
Introduction and Specifications Introduction Introduction The Fluke Biomedical BP Pump 2 Non-Invasive Blood Pressure Simulator and Tester, hereafter known as the “Tester”, is a multi-purpose test instrument for use with oscillometric Non-Invasive Blood Pressure Monitors (NIBPMs). The Tester provides dynamic blood pressure simulations, static calibration, automated leak testing, and pressure relief valve testing.
See manual. of electrical shock. Intertek Electrical Test Laboratory listed. Œ Conforms to European Conforms to relevant Union directives Canadian and U.S. standards. Do not dispose of this product as unsorted municipal waste. Contact Fluke or a qualified recycler for disposal.
Introduction and Specifications General Safety Considerations Warnings and Cautions Users are advised to read the manual carefully, observing all warnings and cautions, before attempting to set up and operate the Tester. A Warning identifies hazardous conditions and actions that could cause bodily harm or death.
Page 18
BP Pump 2 Operators Manual Always connect the system power cord directly to a three-prong receptacle with a functional ground. Never use a two-prong plug adapter to connect primary power to the Tester, thereby disconnecting the utility ground. Disconnect the Tester from the power source before changing the supply voltage.
Introduction and Specifications Instrument Familiarity Instrument Familiarity Figure 1-1 shows the Tester. The top and side panel components are described in Table 1-2 and the number key functions are described in Table 1-3. fas10.eps Figure 1-1. Tester Top and Side Panel Components...
BP Pump 2 Operators Manual Table 1-2. Top and Side Panel Components Label Name Function Home Key Returns the operator to the Main Menu. Makes dynamic assignments based on Soft Keys 1 - 4 the current screen. Allows the operator to perform auto Number (Test and sequences and simulations using Simulations) Keys...
Introduction and Specifications Instrument Familiarity Table 1-3. Number Key Functions Number Name Function Pressurizes a pneumatic system to an operator-defined target pressure up to PRESSURE 400 mmHg (53.3 kPa) and then LEAK measures the loss of pressure over time. Increases the pressure in the pneumatic system until the relief valve PRESSURE on the NIBP monitor opens or until the...
TESTS SEQUENCE SETUP fas31.eps Specifications The following are specifications for the Tester. Please contact your Fluke Biomedical service representative for more information regarding the device specifications. Mains Voltage Range ..............100 - 240 V ac 50/60 Hz, 60 VA Environmental Conditions Operating Temperature ...........15 °C to 40 °C...
BP Pump 2 Operators Manual Accessories The following are accessories for the Tester. To order, contact your Fluke Biomedical equipment dealer and use the Fluke Biomedical part numbers provided. Table 1-4 lists standard accessories shipped with the tester. Table 1-5 lists optional accessories that must be ordered separately.
Chapter 2 Setup, Maintenance, and Support Contents Page Setting up the Tester..............2-3 Printer Output ................2-4 User-Defined Simulations ............2-5 Language .................. 2-6 Units of Measure ..............2-6 Self Test..................2-8 Zero Pressure ................2-8 Enable ECG Signal..............2-9 Maintenance and Support.............
To avoid possible electric shock, burning of the skin, or personal injury, ensure that the external power source is properly rated for the system. 2. Turn the power switch on. When the Tester is powered up, the Main menu appears. Fluke Biomedical BP Bump 2 PERFORM PRESSURE AUTO SETUP...
BP Pump 2 Operators Manual BP Pump 2 Setup UNITS OF RS232 SELF MEASURE SETTINGS TEST MORE* fas38.eps BP Pump 2 Setup ZERO ENABLE PRESSURE MORE* fas39.eps Note These parameters should be configured and the settings saved the first time the Tester is used. They need to be configured only once. Printer Output Printouts are available for auto sequences.
Setup, Maintenance, and Support Setting up the Tester User-Defined Simulations The Tester supports up to nine user-definable blood pressure simulations configured through the Setup menu. To access and modify a user-defined simulation definition, take the following actions: 1. Use the soft keys to follow this menu path: SETUP C USER DEF SIMULATION C Number Key (1-9) C 2.
BP Pump 2 Operators Manual Language Besides the default of English, the Tester can support up to four additional languages. Spanish is currently available; additional languages will be released in the future. To change the language, take the following actions: 1.
Page 31
Setup, Maintenance, and Support Setting up the Tester Blood Pressure Units: mmHg MMHG fas25.eps 2. Select one option and press ENT. The Pressure Measurement Units: screen appears, showing the available options: Pressure Measurement Units: mmHg MMHG CMH2O *MORE fas26.eps 3. Press *MORE to see additional options: Pressure Measurement Units: mmHg INH2O *MORE...
To run a self test, do the following: 1. Use the soft keys to follow the menu path shown below. SETUP C *MORE C SELF TEST The Self Test screen appears: Fluke BP Pump 2 204634 mm Hg 5180) Heise Label: 002422 FW Version: 3.36...
Setup, Maintenance, and Support Setting up the Tester 3. If desired, press HOME to return to the Main menu. Enable ECG Signal The ECG output is enabled or disabled by the operator. If the patient monitor under test does not make use of ECG signals, we recommend that the ECG output be disabled.
BP Pump 2 Operators Manual Maintenance and Support The Tester requires little maintenance or special care; however, it is a calibrated measuring instrument and should be treated as such. The following describes how to maintain the Tester and the method of company support. Avoiding Damage Do not drop the instrument or subject it to any mechanical abuse that could cause a shift in the calibrated settings.
International customers should contact their Fluke Biomedical dealers for service/product support. To obtain the name of a local dealer or service center, contact Fluke Biomedical as indicated under “Return Procedures, Repair and Calibration.” Packing Instructions If repairs are required, return the Tester to the factory or the nearest service center, packed in the original shipping container, using packing materials supplied by Fluke Biomedical.
Page 36
Fluke Biomedical Service. 5. Enclose your return address and Return Authorization Number. 6. Insure the unit for full retail value and ship to the nearest Fluke Biomedical service center. 2-12...
Operation Introduction Introduction The Tester contains a microprocessor that reads and controls the front panel keyboard, the display, serial port, printer port, a diaphragm pump, two solenoid valves, a step motor, a position sensor, and a pressure transducer (Figure 3-1). Filter Check Valve N.C.
BP Pump 2 Operators Manual shape of the pulses are controlled by the microprocessor driving the step motor. The home position of the piston is detected by an optical interrupter. If the optional high-accuracy pressure sensor is installed, it also measures the pressure in the main manifold and is also controlled by the microprocessor.
Page 41
Operation Introduction BP Pump 2 Double Hose: Internal Cuff NON-INVASIVE BLOOD PRESSURE MONITOR ANALYZER NIBP Monitor TESTS AND SIMULATIONS PRESSURE LEAK PATIENT STANDARD BP CONDITIONS ARRHYTHMIAS RESPIRITORY ARTIFACT NEONATE WRIST USER DEFINED fas17.eps Figure 3-3. Connecting Tester to Double-hose NIBP Monitor (Int Cuff) BP Pump 2 Single Hose: External Cuff...
Page 42
BP Pump 2 Operators Manual Double Hose: External Cuff BP Pump 2 NON-INVASIVE BLOOD PRESSURE MONITOR ANALYZER NIBP Monitor TESTS AND SIMULATIONS PRESSURE LEAK PATIENT STANDARD BP CONDITIONS ARRHYTHMIAS Wrap Cuff RESPIRITORY NEONATE ARTIFACT WRIST around Mandrel USER DEFINED Cuff fas19.eps Figure 3-5.
Either method is acceptable, and it is assumed that you will develop your own, most effective way of using the keys. Error Messages If you encounter any repeatable error messages, contact your Fluke Biomedical Service Center. Pressure Tests The following tests assess the integrity and accuracy of the Tester, as well as the...
BP Pump 2 Operators Manual All tests begin from the Main menu: Fluke Biomedical BP Bump 2 PERFORM PRESSURE AUTO SIMULATION TESTS SEQUENCE SETUP fas31.eps After pressing the PRESSURE TESTS soft key to display the Pressure Tests menu, you can access the pressure tests described below:...
Operation Pressure Tests Note Once defined, the Setpoint can be changed in increments of 1 least significant digit (LSD). 3. Press ENT to display the Leak Test screen. Leak Test Leak Rate: _._ _ Setpoint: 250 Cuff: External Measured: 0 CUFF SETUP VENT...
Page 46
BP Pump 2 Operators Manual Note Put the NIBP monitor in “Calibrate” or “Service” mode to close the vent valve, allowing the Tester to inflate the pneumatic system. Refer to the NIBP monitor’s service manual to find the method for entering “Service” mode.
Operation Pressure Tests Note It is recommended that three pressure relief measurements be taken to check for a sticky relief valve. Some NIPB monitors do not allow access to a “Service” mode, rendering it impossible to close a vent valve so that the system can be pressurized by an outside pump.
BP Pump 2 Operators Manual Pressure Source Measured : 0.4 mm Hg Setpoint : 200 mm Hg GAUGE SETUP VENT <START> fas05.eps 4. Press the VENT soft key to release any unwanted pressure in the system before performing the test. This feature vents the system for approximately five seconds and can be repeated as needed to return pressure to zero.
To perform simulations, take the following actions: 1. Power on the NIBP unit. Note Either disable the alarm or be prepared to silence it when it sounds. 2. Use the soft keys, beginning from the Main menu: Fluke Biomedical BP Bump 2 PERFORM PRESSURE AUTO SIMULATION...
BP Pump 2 Operators Manual Select Simulation Type STANDARD PATIENT MORE* CONDITION ARRHYTHMIA fas32.eps 4. Press the MORE soft key on this and the following screens for additional options: Select Simulation Type RESPIRATORY USER ARTIFACTS DEFINED NEONATE MORE* fas33.eps Select Simulation Type WRIST MORE* fas34.eps...
Operation Simulations Standard BP ready ..Preset # 1 120/ 80 (93) Cuff: External 80 BPM 0.68 cc OPTIONS CUFF fas07.eps 2. Press the OPTIONS soft key to scroll through the simulation choices. Parameters for Standard BP simulations are shown in Table 3-2. Table 3-2.
BP Pump 2 Operators Manual Respiratory Artifacts The Respiratory Artifact exhibits a beat-to-beat variation in the blood pressure caused by intra-thoracic pressure. Changes in the intra-thoracic pressure affect filling of the ventricles during diastole. This in turn affects the stroke volume of the heart.
Operation Simulations Neonate The Neonate simulations are provided to test the ability of the NIBP monitors to detect blood pressure on neonatal patients. To perform Neonate simulations, do the following: 1. Access these simulations by pressing the 8 NEONATE key. The Neonate Simulation screen appears.
BP Pump 2 Operators Manual Wrist The Wrist simulations are provided to test wrist cuff NIBP monitors. To perform Wrist simulations, do the following: 1. Access these simulations by pressing the 9 WRIST key. The Wrist Simulation screen appears. Wrist Simulation ready ..
Operation Auto Sequences User-Defined User-Defined simulations and how to create them are described in “Setting Up the Tester.” The following describes how to run such a simulation. 1. Select from any of the previously-defined simulations by pressing the 0 USER DEFINED key. The User Defined screen appears. User Defined ready ..
BP Pump 2 Operators Manual Editing Auto Sequences Use the following steps to edit an auto sequence: 1. Press the AUTO SEQUENCE soft key to access the Auto Sequence menu. Auto Sequence EDIT PRINT fas36.eps 2. Press the EDIT soft key. 3.
Operation Auto Sequences Printing Auto Sequences Use the following steps to print an auto sequence definition. Note Refer to ‘Setting up the Tester” for printer output setup instructions. 1. To determine which tests and simulations are defined in an auto sequence, press the AUTO SEQUENCE soft key to access the Auto Sequence menu.
Operation Auto Sequences Select Auto Sequence: fas42.eps 3. Press a numeric key (1-9), followed by the ENT key. The auto sequence begins to execute, based on the options that were enabled using the EDIT function. Note Tests and simulations not enabled in the executing auto sequence are skipped.
BP Pump 2 Operators Manual 2. If the Leak Rate is unsatisfactory, repair the possible cause of the leak and press START again to repeat the test. 3. Press the NEXT> key to advance to the next test. Only the leak rate displayed when the NEXT>...
Operation Auto Sequences Data Sheet Printout When this step is reached, a printout is generated, as shown in Figure 3-8, which includes a standard header, the results of the previous tests, a place to record data for the Pressure Source test (if enabled), and a list of BP Simulations that will be executed.
Page 64
BP Pump 2 Operators Manual Fluke Biomedical Tester Serial #:123456 Date:________ Time:_______ Serial #: ___________________ Control #: __________________ Mfg: ________________________ Model: ______________________ Location: ___________________ Technician: _________________ Work Order: _________________ Procedure ID: _______________ Gauge: mmHg Leak Test Leak Rate: 5 mmHg/min...
Ansur manages your test procedures by allowing both manual and visual automated test sequences. The software works hand-in-hand with Fluke Biomedical testers and simulators, creating a seamless integration for: Visual inspections...
Page 66
Ansur software by installing a new plug-in. Each plug-in module lets you work only with the options and capabilities you need for the instrument you are testing. The Fluke Part Number for the Ansur BP Pump 2 Plug-in is 2755836.
ECG Interface Optional ECG Interface Adapter The optional ECG Interface Adapter (Fluke Biomedical Part Number 2780512), shown in Figure A-1, is provided to aid in testing NIBP monitors that use ECG signals to assist in the detection of the pressure pulses. The ECG attachment is labeled with the five leads, RA, RL, LA, LL, and V, and is attached to the ECG Interface Port.
Page 70
BP Pump 2 Operators Manual the blood pressure simulations. These ECG signals are available for the Standard BP, Patient Conditions and Arrhythmias simulations.
Appendix B Questions and Answers Introduction The following are questions submitted by customers and answered by members of our technical staff. Blood Pressure Issues Q: Monitor's BP determinations vary. “I connected the Tester to my Critikon DINAMAP monitor and used the preset blood pressure of 120/80 (93) with a pulse rate of 80 beats per minute.
Page 72
BP Pump 2 Operators Manual “The mean difference of the paired measurements of the test system and the comparison system shall be ± 5 mmHg or less with a standard deviation of 8 mmHg or less.” This means that variations in individual readings of 5, 6, or even 10 mmHg are quite normal and do not indicate that either the DINAMAP or the Tester are malfunctioning.
Page 73
Appendices Questions and Answers mean pressures. Further, its compliance is constant over time and is independent of cuff wrapping technique. The compliance of a standard cuff depends on the amount of air it contains. This, in turn, is dependent on what the cuff is wrapped around, and how tightly it is wrapped.
Appendix C Abbreviations Abbreviations The following list includes abbreviations used in this document. ampere ANSI American National Standards Institute Association for the Advancement of Medical AAMI Instrumentation blue (color) beats per minute decibel °C degrees Celsius (centigrade) Contact Quality Monitor digital multimeter EEPROM electrically erasable PROM...
Page 76
BP Pump 2 Operators Manual hertz inch kilo (10 kilogram kilohertz kilohm pound light-emitting diode liquid crystal display meg(a) (10 megahertz megohm meter milli (10 milliampere millimeter millivolt peak-to-peak Return Electrode Monitor second yellow (color) micro (10 µ µA microampere µV microvolt...
Appendix D Computer Control Commands Introduction The following commands are for use by software developers that have highly- developed technical knowledge. Operators should not try to control the Tester using these commands; rather, they should use Ansur test automation software. The Tester acknowledges each valid computer control command that it receives.
Page 78
BP Pump 2 Operators Manual Table D-1. Computer Control Commands (cont.) Returned Description Command String Neonate Cuff Simulation [SIM_NEO35_15] [SIM_NEO60_30] [SIM_NEO80_50] [SIM_NEO100_70] Standard BP Simulation [SIM_STD120_80] [SIM_STD150_100] [SIM_STD200_150] [SIM_STD255_195] [SIM_STD60_30] [SIM_STD80_50] [SIM_STD100_65] Patient Conditions [SIM_HEALTHY] [SIM_WEAK_PULSE] [SIM_MILDEX] [SIM_STRENEX] [SIM_OBESE] [SIM_GERIATRIC] [SIM_TACHYCARDIA] [SIM_BRADYCARDIA]...
Page 79
Appendices Computer Control Commands Table D-1. Computer Control Commands (cont.) Returned Description Command String Arrhythmias [SIM_PAC1] [SIM_PAC2] [SIM_PVC] [SIM_AFIBPVC] Respiratory Artifacts [SIM_SB1] [SIM_SB2] [SIM_SB3] [SIM_CV] Set Units of Blood Pressure [BP_UNITS_KPA] [BP_UNITS_MMHG] Set Units of Pressure [PRESS_UNITS_KPA] [PRESS_UNITS_MMHG] [PRESS_UNITS_CMH2O] [PRESS_UNITS_INH2O] [PRESS_UNITS_PSI]...
Page 80
BP Pump 2 Operators Manual Table D-1. Computer Control Commands (cont.) Returned Description Command String Perform Leak Test at [LEAK,XXX] Returns ACK XXXmmHg 50<=XXX<=400 then Leak Rate. Returns Leak Rate after 60 seconds, followed by CR- LF, or NAK if XXX out of range•...
Page 81
Appendices Computer Control Commands Table D-1. Computer Control Commands (cont.) Returned Description Command String [BP_SERIAL] Returns ACK then the serial number, followed by CR- [BP_VERSION]• ACK then 2780201 Ver X.XX, followed by CR-LF•...