Number Key Functions - Fluke BP Pump 2 Operator's Manual

Nibp simulator and tester
Table of Contents

Advertisement

Number
PRESSURE
1
LEAK
PRESSURE
2
RELIEF
STATIC
3
PRESSURE
4
STANDARD BP
PATIENT
5
CONDITIONS
6
ARRHYTHMIAS
RESPIRATORY
7
ARTIFACT
8
NEONATE
9
WRIST
0
USER DEFINED
Introduction and Specifications
Table 1-3. Number Key Functions
Name
Pressurizes a pneumatic system to an
operator-defined target pressure up to
400 mmHg (53.3 kPa) and then
measures the loss of pressure over
time.
Increases the pressure in the
pneumatic system until the relief valve
on the NIBP monitor opens or until the
Setpoint is reached, whichever occurs
first.
Accessed via the Pressure Gauge
Test, which enables the Tester to
measure static pressure generated by
an external source in the range of 50
to 400 mmHg 6.7 to 53.3 kPa).
Provides seven variations of NIBP
simulations for both arm and wrist
cuffs.
Includes simulations for healthy,
geriatric, and obese patients, a well as
various levels of exercise.
Measures erratic heart rhythms,
including atrial fibrillation and
premature ventricular contraction.
Exhibits a beat-to-beat variation in the
blood pressure caused by intrathoracic
pressure.
Tests the ability of the NIBP monitors
to detect blood pressure on neonatal
patients.
Tests wrist cuff NIBP monitors.
Allows the operator to define blood
pressure simulations.
Instrument Familiarity
Function
1
1-9