Questions And Answers - Fluke BP Pump 2 Operator's Manual

Nibp simulator and tester
Table of Contents

Advertisement

Introduction
The following are questions submitted by customers and answered by
members of our technical staff.
Blood Pressure Issues
Q: Monitor's BP determinations vary.
"I connected the Tester to my Critikon DINAMAP monitor and used the preset
blood pressure of 120/80 (93) with a pulse rate of 80 beats per minute. I
performed three blood pressure determinations with the following results:
Trial #
1
2
3
Why does the blood pressure determined by the DINAMAP vary?"
A: Some variance is normal and acceptable.
The Tester generates a very repeatable simulation. For this simulation, an ideal
NIBP monitor would show a variation of less than 2 mmHg from one
simulation to the next. Most of the variation seen here originates in the
DINAMAP. This is normal and acceptable.
Section 3.4.3 of the ANSI Standard for Electronic or Automated
Sphygmomanometers specifies the required efficacy of the blood pressure
determination:

Questions and Answers

Systolic
123
126
123
Mean
Diastolic
97
93
97
Appendix B
Pulse Rate
82
79
81
81
83
78
B-1