Pressure Gauge Test - Fluke BP Pump 2 Operator's Manual

Nibp simulator and tester
Table of Contents

Advertisement

BP Pump 2
Operators Manual
Pressure Source
GAUGE
4. Press the VENT soft key to release any unwanted pressure in the system
before performing the test. This feature vents the system for approximately
five seconds and can be repeated as needed to return pressure to zero.
5. Press the <START> soft key to make the Tester deliver air to the system. The
Tester pressurizes the system within 10 mmHg of the Setpoint value.
Once the Setpoint has been reached, the Tester will not maintain the
pressure in the system. Therefore, it is recommended that the system be
checked for leaks prior to performing any static pressure tests.
6. Press HOME to return to the Main menu.

Pressure Gauge Test

The Pressure Gauge Test enables the Tester to measure static pressure generated
by an external source in the range of 0 to 400 mmHg (0 to 53.3 kPa).
To measure static pressure, take the following actions:
1. Use the soft keys to follow the menu path shown below.
PRESSURE TESTS C STATIC PRESSURE
The Pressure Gauge screen appears.
Pressure Guage
SOURCE
2. Press the VENT soft key to release any unwanted pressure in the system
before performing the test. This feature vents the system for approximately
five seconds and can be repeated as needed to return pressure to zero.
3-12
Measured : 0.4
Setpoint : 200
SETUP
VENT
Note
Measured: 0
VENT
mm Hg
mm Hg
<START>
0.4 mm Hg
fas05.eps
fas06.eps