User-Defined Simulations

The Tester supports up to nine user-definable blood pressure simulations
configured through the Setup menu.
To access and modify a user-defined simulation definition, take the following
actions:
1. Use the soft keys to follow this menu path:
SETUP C USER DEF SIMULATION C Number Key (1-9) C
ENT
2. Alternatively, you can access the definition from the number keys:
0 USER DEFINED key C OPTIONS (to scroll through the
numbers) C EDIT
With either method, the User Defined Simulation screen appears:
User Defined Simulation 1
Sys: 120
<- FIELD
This screen shows the number of the simulation, along with the following
parameters available for configuration. Their valid ranges are as follows:
Sys: Systolic, 20 – 250 mmHg
Dia: Diastolic, 10 – 200 mmHg
Pulse: Pulse Volume, 0.1 cc – 2.4 cc in increments of 0.1 cc
HR: Heart Rate, 30 – 250 bpm
The systolic and diastolic settings are interdependent. The diastolic must
always be below the systolic. The pulse volume and the heart rate are also
interdependent. The maximum pulse volume cannot be achieved at the
maximum heart rate.
3. Press the soft key <- FIELD or FIELD -> to scroll through options,
making changes to each field by entering numbers from the number keys.
4. Pres ENT to return to the SETUP menu.
5. If desired, press the HOME button to return to the Main menu.
Setup, Maintenance, and Support
Dia: 80
Pulse: 0.70cc
FIELD ->
Setting up the Tester
HR: 75
2
fas23.eps
2-5