Set The N-Cycle Burst - Owon AG4151 User Manual

Arbitrary waveform generator
Hide thumbs Also See for AG4151:
Table of Contents

Advertisement

Set the N-Cycle Burst

Load
Start Phase
N Cycle
Waveform
Source
Waveform
Figure 5-20: The User Interface of N-Cycle Burst
(1) When the output signal is Sine, Square, Ramp, Pulse or Arbitrary waveform, press
Burst function button to enter the Burst mode.
(2) Press
,
Sine for instance, press
change the parameters, please refer to "To set signals" on P11. Press
again to return to the Burst mode interface.
(3) Press F4 to switch to N Cycle.
(4) Press F1 to select Burst Period, set it to the desired value.
(5) Press F3 to select Start Phase, define the Start and the Stop Point in a waveform.
The phase varies from -360° to +360°. For an Arbitrary Waveform, 0° is the first
waveform point.
(6) Press F2 to select Delay, set the Time Delay between the Trigger Input and the
Start of the N-Cycle Burst. The minimum delay is a function of the specific burst
period, and should always be greater than 0s.
(7) Press F5 to select NextPage, press it again to enter the next page.
(8) Press F1 to select Cycles or Infinite. Set the number of Waveform Cycle in an
N-Cycle (from 1 to 50000). If you choose Infinite, then a continuous waveform will
be generated which will not stop until a trigger event happens ( Trigger button is
pressed).
Note:
If needed, Burst Period will increase to cater to the specific number of cycles.
For an infinite-cycle Burst, External or Manual Trigger is needed to activate burst.
(9) Press F2 to select the source. Internal means using the internal source. External
means using the Ext Trig/FSK/Burst connector in the rear panel to input the external
signal. Manual means using the external source, set the start and stop time by hand.
5.Front Panel Operation
Current Parameter
,
,
or
to display the waveform and parameters. You can
28
Burst Period
button to choose the waveform. Take
Cycles
Delay
button

Advertisement

Table of Contents
loading