Hanna Instruments HI 710 Series Instruction Manual page 9

Process, panel-mounted, microprocessor-based, conductivity and tds controllers
Hide thumbs Also See for HI 710 Series:
Table of Contents

Advertisement

Code
60 Current day
61 Current month
62 Current year
63 Current time
71 Baud rate (RS485)
72 Cleaning timer
73 Initial cleaning day
74 Initial cleaning month
75 Initial cleaning year
76 Initial cleaning time
77 Cleaning ON interval
90 Display selftest
91 Keyboard selftest
92 EEPROM selftest
93 Relays and LEDs selftest
94 Watchdog selftest
99 Unlock password
Note
The process controller automatically checks to ensure that
the entered data matches other related variables. If a wrong
configuration is entered, "ERROR" blinks on the LCD to prompt
the user. The correct configurations are the following:
If M1= / 0 then S1 HA, S1 LA;
If M2= / 0 then S2 HA, S2 LA;
If M1= 1 then S1-H1 LA;
If M1= 2 then S1+H1 HA;
16
Valid Values
Default
PW
01 to 31
from RTC
no
01 to 12
from RTC
no
1998 to 9999
from RTC
no
00:00 to 23:59
from RTC
no
1200, 2400, 4800, 9600
9600
no
0 to 19999 days
0
no
01 to 31
01
no
01 to 12
01
no
1998 to 9999
1998
no
00:00 to 23:59
00:00
no
0 to 19999 minutes
0
no
0: off
0
yes
1: on
0: off
0
yes
1: on
0: off
0
yes
1: on
0: off
0
yes
1: on
0: off
0
yes
1: on
0000 to 9999
0000
yes
If M1= 3 then S1+D1 HA;
If M1= 4 then S1-D1 LA;
If M2= 1 then S2-H2 LA;
If M2= 2 then S2+H2 HA;
If M2= 3 then S2+D2 HA;
If M2= 4 then S2-D2 LA;
If M1= 1 and M2 = 2
then S1-H1 S2+H2, S2 LA, HA S1;
If M1 = 2 and M2 = 1
then S2-H2 S1+H1, S1 LA, HA S2;
If M1 = 3 and M2 = 2
then S1 S2+H2, S2 LA, HA S1+D1;
If M1 = 2 and M2 = 3
then S1+H1 S2, S1 LA, HA S2+D2;
If M1 = 4 and M2 = 1
then S1 S2–H2, S1–D1 LA, HA S2;
If M1 = 1 and M2 = 4
then S1–H1 S2, S2–D2 LA, HA S1;
If M1 = 3 and M2 = 4
then S1 S2, S2–D2 LA, HA S1+D1;
If M1 = 4 and M2 = 3
then S2 S1, S1–D1 LA, HA S2+D2;
where the minimum deviation (D1 or D2) is 0.5% of the maxi-
mum range value.
Note
When a wrong setup value is confirmed,
the controller does not skip to the next setup
item but remains in the current item dis-
playing a flashing "ERROR" indicator un-
til the parameter value is changed by the
user (the same is also true for the setup code selection).
Note
In some circumstances, the user cannot succeed in setting a
parameter to a desired value if the related parameters are not
changed beforehand; e.g. to set a EC high setpoint to 10.0
mS the high alarm must be set to a value greater than 10.0
mS first.
17

Hide quick links:

Advertisement

Table of Contents
loading