Hanna Instruments HI8510 Instruction Manual

Hanna Instruments HI8510 Instruction Manual

Panel mounted ph and orp indicators and controllers
Hide thumbs Also See for HI8510:

Advertisement

Instruction Manual
HI8510 • HI8512
HI8710 • HI8711
HI8720
Panel Mounted
pH and ORP
Indicators and
Controllers
w w w . h a n n a i n s t . c o m
1

Advertisement

Table of Contents
loading

Summary of Contents for Hanna Instruments HI8510

  • Page 1 Instruction Manual HI8510 • HI8512 HI8710 • HI8711 HI8720 Panel Mounted pH and ORP Indicators and Controllers w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instru‑ ment is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid.
  • Page 3: Preliminary Examination

    GENERAL DESCRIPTION HI8510 and HI8512 pH and ORP panel‑mounted indicators, and HI8710, HI8711 and HI8720 pH and ORP controllers, are ideal for process control monitoring in a wide range of industrial applications.
  • Page 4: Mechanical Dimensions

    MECHANICAL DIMENSIONS The meters are provided with a black anodized aluminum body, front and back panels in shockproof ABS plastic and a transparent splash‑proof front cover. Front view of the panel-mounted unit The dimensions show the cutout size for the installation. Side view of the panel-mounted unit Adjustable location brackets (supplied with the meter) allow the indicator to slide into the cutout and will hold the unit securely in...
  • Page 5: Functional Description Hi8510

    FUNCTIONAL DESCRIPTION HI8510 FRONT PANEL Keypad SENSOR TEST To display the mV reading of the electrode and, therefore, verify its working condition pH7 TEST To verify the internal circuit of the meter in terms of Offset compensation pH4 TEST To verify the amplifier circuit of the meter...
  • Page 6 REAR PANEL HI8510 1. BNC socket for pH electrode 2. Input from amplified electrode 3. Connections for Pt100 temperature sensor 4. Power supply terminals 5. Fuse holder 6. Recorder output terminals 7. Connection to the transmiter 8. Power for amplified electrode 9.
  • Page 7: Functional Description Hi8512

    FUNCTIONAL DESCRIPTION HI8512 FRONT PANEL Keypad 0 mV TEST To verify the instrument calibration at 0 mV 1000 mV TEST To verify the slope at 1000 mV Note: Each time a key is pressed the corresponding LED is turned Trimmers For ORP calibration...
  • Page 8 REAR PANEL HI8512 1. BNC socket for ORP electrode 2. Input from amplified electrode 3. Power supply terminals 4. Fuse holder 5. Recorder output terminals 6. Connection to the transmitter 7. Power for amplified electrode 8. Connection for matching pin 9.
  • Page 9: Functional Description Hi8710

    FUNCTIONAL DESCRIPTION HI8710 FRONT PANEL Keypad To set the pH dosage limit MEASURE To enter measurement mode and to enable diagnostic tests SENSOR TEST To display electrode mV reading and verify its working condition ∆ALARM To display & set alarm tolerance pH7 TEST To verify Offset compensation pH4 TEST...
  • Page 10 LEDs ACID Shows that acid dosage is (blinking) active ALK. Shows that alkaline dosage is blinking) active ∆ALARM Indicates an active alarm (blinking) DOSAGE MODE SWITCH Blinks when the dosing switch is in OFF or ON position. SWITCH OFF/AUTO/ON To set the dosing mode: •...
  • Page 11 REAR PANEL HI8710 1. BNC socket for pH electrode 2. Input from amplified electrode 3. Connections for Pt100 temperature sensor 4. Connections for dosing pump 5. Reduc/Oxid dosage consent terminals 6. Alarm contacts 7. Power supply terminals 8. Fuse holder 9.
  • Page 12: Functional Description Hi8711

    FUNCTIONAL DESCRIPTION HI8711 FRONT PANEL Keypad ALK. SET To set the working point of basic dos‑ ACID SET To set the working point of acid dosage MEASURE To enter measurement mode and to enable diagnostic tests To display electrode mV reading and, SENSOR TEST therefore, verify its working condition ∆ALARM...
  • Page 13 Trimmers SLOPE For Slope calibration OFFSET For Offset calibration ∆ALARM To set the tolerance of the alarm ALK. SET To coarsely adjust alkaline setpoint COARSE To finely adjust alkaline setpoint FINE ACID SET To coarsely adjust acid setpoint COARSE To finely adjust acid setpoint FINE LEDs ALK.
  • Page 14 REAR PANEL HI8711 1. BNC socket for pH electrode 2. Input from amplified electrode 3. Connections for Pt100 temperature sensor 4. Connections for dosing pump for acid 5. Connections for dosing pump for base 6. Alarm contacts 7. Power supply terminals 8.
  • Page 15: Functional Description Hi8720

    FUNCTIONAL DESCRIPTION HI8720 FRONT PANEL Keypad To set the working point of ORP dosage MEASURE To enter measurement mode and to enable diagnostic tests ∆ALARM To display and set the alarm tolerance 0 mV TEST To verify the instrument calibration at 0 mV 1000 mV TEST To verify the slope at 1000mV Note: Each time a key is pressed the corresponding LED is turned ON...
  • Page 16 Trimmers For ORP calibration ∆ALARM To display and set the alarm tolerance SET/COARSE To coarsely adjust the setpoint SET/FINE To finely adjust the setpoint LEDs OXID Show that the oxidant dosage is active (Blinking) RED. Show that the reductant dosage is active (Blinking) ∆ALARM Indicate an active alarm...
  • Page 17 REAR PANEL HI8720 1. BNC socket for ORP electrode 2. Input from amplified electrode 3. Oxid/Reduc dosage consent terminals 4. Connections for dosing pump 5. Alarm contacts 6. Power supply terminals 7. Fuse holder 8. OXID/RED. dosage selection terminals 9. Recorder output contacts 10.
  • Page 18: Specifications

    SPECIFICATIONS HI8510 INPUT Electrode Transmitter RANGE 0.00 to 14.00 pH RESOLUTION 0.01 pH ACCURACY ±0.02 pH (0 to 100 °C) ±0.5% (@25 °C/77 °F) ±0.05 pH (‑20 to 0 °C) High impedance 10 Ohm; INPUTS Reference and Matching pin inputs...
  • Page 19 HI8710 INPUT Electrode Transmitter RANGE 0.00 to 14.00 pH RESOLUTION 0.01 pH ACCURACY ±0.02 pH (0 to 100 °C) ±0.5% (@25 °C/77 °F) ±0.05 pH (‑20 to 0 °C) High impedance 10 Ohm; INPUTS Reference and Matching pin inputs 4 to 20 mA are available POWER OUTPUT ±5 Vcc;...
  • Page 20 RECORDER OUTPUT 0‑20 mA or 4‑20 mA (isolated) SETPOINT RELAY 1, isolated, 2 A, Max 240 V, resistive load, 1000000 strokes (not fuse protected) SETPOINT RANGE 0.00 to 14.00 pH ALARM RELAY 1, isolated, 2 A, Max 240 V, resistive load, 1000000 strokes (not fuse protected) ∆ALARM RANGE 0.2 to 3.00 pH DOSING CONTROL...
  • Page 21: Initial Preparation

    INITIAL PREPARATION • Connect a 3‑wire cable to the power supply terminal according to the voltage level as indicated, and pay particular attention to the correct line, earth and neutral connections. • For BNC electrodes, connect the electrode to the BNC plug on the rear panel. The instruments are equiped with differential input.
  • Page 22 Connect the “+” wire of the recorder to the terminal 1 on the instrument and the other wire (common) to terminal 2 for 4‑20 mA recorder output or to terminal 3 for 0‑20 mA recorder output. Note: Only one recorder output connection is possible. In order to avoid malfunction leave the unused terminal unconnected.
  • Page 23 • ALK. contact (HI8711) (see picture, terminals 3 and 4): these contacts are used to connect ALK. the dosing pump for base, and act as a ALARM switch for the power to the drive. • CONSENT contacts (HI8710 and HI8720, see picture, terminals 3 and 4): these contacts (max.
  • Page 24: Operational Guide

    OPERATIONAL GUIDE All instrument settings are made via front panel keys and trimmers. When a key is pressed, the corresponding LED lights up to show the operating function. If the LED blinks before pressing the key, it will change the blinking style. If using an asigned electrode, make sure that the meter is calibrated before starting any operation (see “Calibration”...
  • Page 25 Using a small screwdriver adjust the COARSE and FINE trimmers to display the desired set value. SET POINTS (HI8711) To set the working point for alkaline dosage, press the ALK. SET key and the display will show the set value for alkaline dosage. The ALK. SET LED will turn ON if alkaline dosage is deactivated or change blinking style.
  • Page 26 Note: The ALK. and the ACID setpoints can be set on the entire range 0.00 pH to 14.00 pH using COARSE and FINE trimmers. In order to avoid erroneous situations the ALK. setpoint value should not exceed the ACID setpoint value. ALARMS (HI8710, HI8711 and HI8720) To set the alarm tolerance, press ∆ALARM key and the display will show the current value.
  • Page 27 The actual pH or ORP value of the test solution is displayed. When acid dosage is active, the ACID LED lights DOSING up, while during alkaline dosage, the ALK. LED SLOPE ALK. OFF/AUTO/ON OFFSET ACID turns on (HI8710 only). ∆ALARM MEASURE When oxidant dosage is active, the OXID LED lights up, while during reductant dosage, the...
  • Page 28: Ph Calibration

    Two DOSAGE MODE switches are designed for HI8711, one for ACID and the other for ALK. channels. If for any reason one of them is in OFF position the corresponding dosing relay is deactivated. The alarm relay will be activated only in accordance with the other channel. The ALARM LED will work as in AUTO mode.
  • Page 29 Rinse pH electrode and the reference thermometer thoroughly with pH4.01 rinsing solution, then immerse them in pH4.01 (HI7004) or pH10.01 (HI7010) buffer solution. Note: For accurate readings, use pH4.01 if you are going to measure acid samples or pH10.01 for alkaline measurements. Shake briefly and wait one minute before adjusting the SLOPE trimmer to display the ph value of the buffer solution, i.e.
  • Page 30: Ph Values At Various Temperature

    pH VALUES AT VARIOUS TEMPERATURE Temperature has an effect on the pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. Please refer to the following chart to perform the pH calibration: TEMP pH BUFFERS ºC ºF...
  • Page 31: Ph Diagnostic Tests

    DIAGNOSTIC TESTS HI8510, HI8710 and HI8711 are provided with autodiagnostic func‑ tions that allow to check and troubleshoot any malfunctioning. The functions are made via front panel keys to isolate the cause of malfunction whether it is due to pH electrode contamination, internal offset circuit or amplifier circuit.
  • Page 32 C) Amplifier Circuit Test Press the pH4 TEST key and the display should show a value within the 3.30 to 4.30 pH range, to verify the amplifier circuit of the meter. The corresponding LED turns on.
  • Page 33: Orp Diagnostic Tests

    ORP DIAGNOSTIC TESTS HI8512 and HI8720 are ORP controllers provided with autodiagnostic functions that allow to check and troubleshoot any malfunctioning. The functions are made via front panel keys to isolate the cause of malfunction. For HI8720 only, press MEASURE key before proceeding with the following tests.
  • Page 34: Led Indication

    LED INDICATION All LEDs above the keys indicate the state of each function, whether it is active or the display is indicating the mode. For HI8711 only Each LED can be in one of the following states: A) Light on The mode is displayed on the LCD but is not active, e.g.
  • Page 35: Taking Redox Measurements

    TAKING REDOX MEASUREMENTS Redox measurements allow the quantification of the solution oxidizing/ reducing power, and are commonly expressed in mV. Oxidation may be defined as the process during which a molecule (or an ion) loses electrons and reduction as the process by which electrons are gained.
  • Page 36 Oxidizing pre‑treatment: immerse the electrode for some minutes in HI7092 solution. If no pre‑treatment is performed, the electrode will have long response times. If working with refillable electrodes, always check the internal electrolyte level and refillwith HI7071 solution, if necessary (the level must be at least 2.5 cm below the filling hole).
  • Page 37: Electrode Maintenance

    ELECTRODE MAINTENENCE PREPARATION PROCEDURE Remove the protective cap. DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is nor‑ mal with electrodes and they will disappear when rinsed with water. During transport tiny air bubbles may form inside the glass bulb, and the electrode cannot function properly under these conditions.
  • Page 38 Refill the electrode with fresh electrolyte solution (HI7071 for single junction electrodes or HI7082 for double junction). Allow the electrode to stand upright for 1 hour. Follow the above “Storage Procedure”. AmpHel is a registered Trademark of “Hanna Instruments” ®...
  • Page 39 Soak in Hanna Instruments HI7074 inorganic clean‑ ing solution for 15 minutes. • Oil/grease Rinse with Hanna Instruments HI7077 oil & fat cleaning solution. IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled water, drain and refill the reference chamber with fresh electrolyte, (not necessary for gel‑filled electrodes)
  • Page 40: Suggested Installations

    ® signal, drastically reducing susceptibility to noise and drift. The sealed electrode body can stand a moisture up to 100% RH without any effect on the signal. AmpHel is a registered Trademark of “Hanna Instruments” ®...
  • Page 41 If the needed installation distance is greater than 50 m (165’), it is necessary the use of a transmitter. Hanna Instruments offers a full line of pH and ORP transmitters with ® or without display. AmpHel is a registered Trademark of “Hanna Instruments” ®...
  • Page 42: Accessories

    ACCESSORIES pH CALIBRATION SOLUTIONS HI7004M pH4.01 buffer solution, 230 mL HI7004L pH4.01 buffer solution, 500 mL HI7006M pH6.86 buffer solution, 230 mL HI7006L pH6.86 buffer solution, 500 mL HI7007M pH7.01 buffer solution, 230 mL HI7007L pH7.01 buffer solution, 500 mL HI7009M pH9.18 buffer solution, 230 mL HI7009L...
  • Page 43 pH ELECTRODES HI1090T Screwcap PG13.5 connector, double junction, glass body HI1110S Screw connector, single junction, glass body HI1130B/3 BNC connector, 3 m (9.9’) cable, single junction, glass body HI1110T Screwcap PG13.5 connector, double junction, glass body HI1114S Screw connector, double junction, plastic body HI1134B/3 BNC connector, 3 m (9.9’) cable, double junction plastic body...
  • Page 44 HI1210T Screwcap PG13.5 connector, double junction, plastic body HI1910B BNC connector, 1 m (3.3’) cable, double junction, plastic body, built‑in amplifier HI1911B BNC connector, 1 m (3.3’) cable, double junction, plastic body, built‑in amplifier HI1912B BNC connector, 1 m (3.3’) cable, double junction, plastic body, built‑in amplifier HI1912B/5 BNC connector, 5 m (16.5’) cable, double junction,...
  • Page 45 ORP ELECTRODES HI2930B/5 BNC connector, 5 m (16.5’) cable, Pt, plastic body, built‑in amplifier HI3110S Screw connector, Pt, glass body HI3130B/3 BNC connector, 3 m (9.9’) cable, Pt, glass body HI3110T Screwcap PG13.5 connector, Pt, glass body HI3115S Screw‑type connector, side‑arm, Pt, glass body HI3135B/3 BNC connector, 3 m (9.9’) cable, side‑arm, Pt, glass body...
  • Page 46 HI3932B/5 BNC connector, 5 m (16.5’) cable, Pt, plastic body, built‑in amplifier HI4110S Screw connector, Au, glass body HI4130B/3 BNC connector, 3 m (9.9’) cable, Au, glass body HI4932B/5 BNC connector, 5 m (16.5’) cable, Au, plastic body, built‑in amplifier OTHER ACCESSORIES ChecktempC thermometer with penetration probe HI98501...
  • Page 47 24 VAC or 60 VDC. To avoid damage or burns, do not perform any measurement in microwave ovens. Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
  • Page 48 Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E‑mail tech@hannainst.com www.hannainst.com Printed in EUROPE (ROMANIA) MAN8711 10/17...