ANALOG OUTPUT CALIBRATION
In the meters where the analog output is available, this fea-
ture is factory calibrated through software. The user may also
perform these calibration procedures.
IMPORTANT
It is recommended to perform the out-
put calibration at least once a year.
Calibration should only be performed
after 10 minutes from power up.
• With a multimeter or an HI 931002 connect the common
port to the ground output and the second port to the cur-
rent output or voltage output (depending on which
parameter is being calibrated).
• Press and hold in sequence CFM first, then
CAL to enter the Analog Output Calibration mode.
• Execute the password procedure.
• The primary display will show the current selected param-
eter blinking. Use the
below) for the desired parameter displayed on the second-
ary display (e.g. 4-20 mA).
• Press CFM to confirm the selected parameter that will stop
blinking on the primary display. The secondary display shows
the HI 931002 or multimeter input value as lower limit of
the interval.
• Use the
HI 931002 or multimeter out-
put correspond with the meter's
42
and then
to select the code (0-5 see chart
or
to make the
value shown on the secondary display (e.g. 4).
• Wait for approximately 30 seconds (until the reading of the
calibrator is stable).
• Press CFM to confirm. The meter will switch to the second
calibration point. Repeat the above procedure.
• After the desired readings are obtained, press CFM and
the meter will skip back to normal operational mode.
Note
When adjusting values using the
to allow for sufficient response time (up to 30 seconds)
The table below lists the values of output codes along with
the calibration point values (which are the analog output
minimum and the analog output maximum) as indicated on
the display.
The secondary display indicates the current calibration point
value, while primary display indicates the current calibration
type.
OUTPUT
CALIBRATION
TYPE
CODE
0-1 mA
0
0-20 mA
1
4-20 mA
2
0-5 Vdc
3
1-5 Vdc
4
0-10 Vdc
5
or
keys it is important
CALIBRATION
CALIBRATION
POINT 1
POINT 2
0 mA
1 mA
0 mA
20 mA
4 mA
20 mA
0 Vdc
5 Vdc
1 Vdc
5 Vdc
0 Vdc
10 Vdc
43