Hanna Instruments HI 8510 Instruction Manual

Hanna Instruments HI 8510 Instruction Manual

Panel - mounted orp - ph indicators and controllers
Hide thumbs Also See for HI 8510:

Advertisement

Instruction Manual
HI 8510 - HI 8512
HI 8710 - HI 8711
HI 8720 - HI 931500
HI 931501 - HI 932500
Panel - Mounted
ORP - pH Indicators
and Controllers
HI 8720E
ORP
mV
DOSAGE
SLOPE
OXID.
REDUC.
∆AL
MEA
SET
SURE
COARSE
SET
∆AL
FINE
0 mV
500 mV
TEST
TEST
These Instruments are in Compliance with the CE Directives
h t t p : / / w w w . h a n n a i n s t . c o m

Advertisement

Table of Contents
loading

Summary of Contents for Hanna Instruments HI 8510

  • Page 1 Instruction Manual HI 8510 - HI 8512 HI 8710 - HI 8711 HI 8720 - HI 931500 HI 931501 - HI 932500 Panel - Mounted ORP - pH Indicators and Controllers HI 8720E DOSAGE SLOPE OXID. REDUC. ∆AL SURE COARSE ∆AL...
  • Page 2: Table Of Contents

    Functional Description HI 931500 .... 22 Functional Description HI 931501 .... 24 Functional Description HI 932500 .... 27 Specifications HI 8510 and HI 8512 ..29 Specifications HI 8710 and HI 8711 ..30 Specifications HI 8720 and HI 931500 ..31 Specifications HI 931501 and HI 932500 .
  • Page 3: Preliminary Examination

    GENERAL DESCRIPTION HI 8510, HI 8512, HI 8710, HI 8711, HI 8720, HI 931500, HI 931501 and HI 932500 are pH- ORP panel-mounted indicators and control- lers designed for simplicity of use in a wide range of industrial process applications.
  • Page 4 Using the pH indicators in conjunction with a 4-20mA output pH transmitter HI 8614 or HI 8614L (with LCD display) and using the ORP indicators in conjunction with a 4-20 mA output ORP transmitter HI 8615 or HI 8615L (with LCD display) will assure you of a strong, interference-free signal at distances up to 300 m (1000 ft).
  • Page 5: Mechanical Dimensions

    MECHANICAL DIMENSIONS The meters have a DIN 43 700 casing in black anodized aluminum. Front and back of the instruments are supplied with shockproof ABS plastic and a transparent protective cover for the front panel. Front view of the panel-mounted unit. 141mm 144mm 5.55"...
  • Page 6: Functional Description Hi 8510

    FUNCTIONAL DESCRIPTION HI 8510 pH INDICATOR FRONT PANEL HI 8510E SLOPE ∆ SENSOR TEST pH 7 pH 4 TEST TEST Keypad SENSOR TEST To display the mV response of the electrode in order to verify its working condition pH 7 TEST...
  • Page 7 REAR PANEL HI 8510E ELECTRODE INPUT PT100 mA OUT FUSE FUSE 1. Fuse Holder 2. Power supply 3. Recorder output 4. Connections for PT100 temperature sensor 5. BNC socket for pH electrode. Unplug the instrument from the power supply before replacing the fuse.
  • Page 8 REAR PANEL HI 8510T 1. Fuse Holder 2. Power supply 3. Recorder output 4. Connections to the transmitter. Unplug the instrument from the power supply before replacing the fuse.
  • Page 9: Functional Description Hi 8512

    FUNCTIONAL DESCRIPTION HI 8512 ORP INDICATOR FRONT PANEL HI 8512E SLOPE 500 mV 0 mV TEST TEST Keypad 0 mV TEST To verify the instrument cali- bration at point 0 500 mV TEST To verify the slope at point 500 mV Trimmers SLOPE For Slope calibration...
  • Page 10 REAR PANEL HI 8512E ELECTRODE INPUT mA OUT FUSE FUSE 1. Fuse Holder 2. Power supply 3. Recorder output 4. BNC socket for ORP electrode. Unplug the instrument from the power supply before replacing the fuse.
  • Page 11 REAR PANEL HI 8512T 1. Fuse Holder 2. Power supply 3. Recorder output 4. Connections to the transmitter. Unplug the instrument from the power supply before replacing the fuse.
  • Page 12: Functional Description Hi 8710

    FUNCTIONAL DESCRIPTION HI 8710 pH CONTROLLER WITH ALARM FRONT PANEL HI 8710E SLOPE ACID BASE ∆ ∆AL SURE COARSE SENSOR ∆AL TEST FINE pH 7 pH 4 TEST TEST Keypad To set the working point of pH dosage MEASURE To set HI 8710 on measure- ment mode and to enable the diagnostic tests SENSOR TEST To display the mV response...
  • Page 13 Trimmers ∆ ∆ ∆ ∆ ∆ O For Offset calibration SLOPE For Slope calibration ∆ ∆ ∆ ∆ ∆ AL To set the tolerance of the alarm SET/COARSE To coarsely adjust the set point SET/FINE To finely adjust the set point Leds ACID To show the acid dosage is...
  • Page 14 REAR PANEL HI 8710E ELECTRODE INPUT PT100 mA OUT OPEN: ACID SHORT: BASE CONSENT FUSE ALARM FUSE 1. Fuse Holder 2. Power supply 3. Acid/Basic dosage selection 4. Reductant or oxidant dosage consent 5. Connections for dosing pump 6. Alarm contacts 7.
  • Page 15 REAR PANEL HI 8710T 1. Fuse Holder 2. Power supply 3. Acid/Basic dosage selection 4. Reductant or oxidant dosage consent 5. Connections for dosing pump 6. Alarm contacts 7. Recorder output 8. Connections to the transmitter. Unplug the instrument from the power supply before replacing the fuse.
  • Page 16: Functional Description Hi 8711

    FUNCTIONAL DESCRIPTION HI 8711 DUAL OUTPUT pH CONTROLLER FRONT PANEL HI 8711E SLOPE ACID BASE ∆ ∆AL SURE COARSE ACID SET SENSOR ∆AL TEST FINE COARSE pH 7 pH 4 BASE SET TEST TEST FINE HI 8711E Keypad ACID SET To set the working point of acid dosage BASE SET...
  • Page 17 pH 7 TEST To verify the internal cir- cuit of the meter in terms of Offset compensation pH 4 TEST To verify the amplifier cir- cuit of the meter Trimmers ∆ ∆ ∆ ∆ ∆ O For Offset calibration SLOPE For Slope calibration ∆...
  • Page 18 REAR PANEL HI 8711E ELECTRODE INPUT PT100 mA OUT BASE ACID FUSE ALARM FUSE 1. Fuse Holder 2. Power supply 3. Connections for dosing pump for acid 4. Connections for dosing pump for base 5. Alarm contacts 6. Recorder output 7.
  • Page 19 REAR PANEL HI 8711T 1. Fuse Holder 2. Power supply 3. Connections for dosing pump for acid 4. Connections for dosing pump for base 5. Alarm contacts 6. Recorder output 7. Connections to the transmitter. Unplug the instrument from the power supply before replacing the fuse.
  • Page 20: Functional Description Hi 8720

    FUNCTIONAL DESCRIPTION HI 8720 ORP CONTROLLER FRONT PANEL HI 8720E DOSAGE SLOPE OXID. REDUC. ∆AL SURE COARSE ∆AL FINE 500 mV 0 mV TEST TEST Keypad To set the working point of ORP dosage MEASURE To set HI 8720 on measurement mode and to enable the diag- nostic tests ∆...
  • Page 21 Trimmers SLOPE For Slope calibration ∆ ∆ ∆ ∆ ∆ AL To display and set the tolerance of the alarm points SET/COARSE To coarsely adjust the set point SET/FINE To finely adjust the set point Leds OXID To show the oxidant dosage is active REDUC To show the reductant dosage is...
  • Page 22 REAR PANEL HI 8720E ELECTRODE INPUT CONSENT mA OUT OPEN: OXD SHORT: RDX FUSE ALARM FUSE 1. Fuse Holder 2. Power supply 3. Oxidant/Reductant dosage selection 4. Oxidant or reductant dosage consent 5. Connections for dosing pump 6. Alarm contacts 7.
  • Page 23 REAR PANEL HI 8720T 1. Fuse Holder 2. Power supply 3. Oxidant/Reductant dosage selection 4. Oxidant or Reductant dosage consent 5. Connections for dosing pump 6. Alarm contacts 7. Recorder output 8. Connections to the transmitter. Unplug the instrument from the power supply before replacing the fuse.
  • Page 24: Functional Description Hi 931500

    FUNCTIONAL DESCRIPTION HI 931500 SINGLE OUTPUT pH CONTROLLER FRONT PANEL ins rum n s DOSAGE SLOPE ACID BASE ∆O SURE COARSE FINE HI 931500 pH Process Controller Keypad To set the working point of pH dosage MEASURE To set HI 931500 on measure- ment mode Trimmers ∆...
  • Page 25 REAR PANEL HI 931500 ELECTRODE INPUT PT100 mA OUT OPEN : ACID SHORT : BASE CONSENT FUSE FUSE 1. Fuse Holder 2. Power supply 3. Reductant or oxidant dosage consent 4. Acid/Basic dosage selection 5. Connections for Dosing Pumps 6. Recorder output 7.
  • Page 26: Functional Description Hi 931501

    FUNCTIONAL DESCRIPTION HI 931501 DUAL OUTPUT pH CONTROLLER FRONT PANEL ins rum n s DOSAGE SLOPE ACID BASE ∆O SURE COARSE ACID SET FINE COARSE HI 931501 BASE SET pH Process Controller FINE Keypad ACID SET To set the working point of acid dosage BASE SET To set the working point of basic dosage...
  • Page 27 ACID SET/FINE To finely adjust the acid set point BASE SET/COARSE To coarsely adjust the basic set point BASE SET/FINE To finely adjust the ba- sic set point Leds ACID SET (Blinking) To show the acid dos- age is active BASE SET (Blinking) To show the basic dosage is active MEASURE To show the meter is on mea-...
  • Page 28 REAR PANEL HI 931501 ELECTRODE INPUT PT100 mA OUT BASE ACID FUSE FUSE 1. Fuse Holder 2. Power supply 3. Connections for dosing pump for acid 4. Connections for dosing pump for base 5. Recorder output 6. Connections for PT100 temperature sensor 7.
  • Page 29: Functional Description Hi 932500

    FUNCTIONAL DESCRIPTION HI 932500 ORP CONTROLLER FRONT PANEL DOSAGE SLOPE OXID. REDUC. SURE COARSE FINE HI 932500 ORP Process Controller Keypad To set the working point of ORP dosage MEASURE To set HI 932500 on measure- ment mode Trimmers SLOPE For Slope calibration SET/COARSE To coarsely adjust the ORP set point...
  • Page 30 REAR PANEL HI 932500 ELECTRODE INPUT CONSENT mA OUT OPEN: OXD SHORT: RDX FUSE FUSE 1. Fuse Holder 2. Power supply 3. Oxidant or reductant dosage consent 4. Oxidant/Reductant dosage selection 5. Connections for dosing pump 6. Recorder output 7. BNC socket for ORP electrode. Unplug the instrument from the power supply before replacing the fuse.
  • Page 31: Specifications Hi 8510 And Hi 8512

    SPECIFICATIONS HI 8510 HI 8510E HI 8510T Range 0.00 to 14.00 pH Resolution 0.01 pH Accuracy ±0.02 pH ±0.5% Typical EMC ±0.1 pH ±0.1 pH Deviation ±0.2 mA ±0.2 mA Installation Cat. Input 4 to 20 mA Offset: ±2 pH ∆O trimmer...
  • Page 32: Specifications Hi 8710 And Hi 8711

    SPECIFICATIONS HI 8710 HI 8710E HI 8710T Range 0.00 to 14.00 pH Resolution 0.01 pH Accuracy ±0.02 pH ±0.5% Typical EMC ±0.1 pH ±0.1 pH Deviation ±0.2 mA ±0.2 mA Installation Cat. Input 4 to 20 mA Offset: ±2 pH ∆O trimmer Calibration Slope: 80 to 110% Slope trimmer Temperature...
  • Page 33: Specifications Hi 8720 And Hi 931500

    SPECIFICATIONS HI 8720 HI 8720E HI 8720T Range -1000 to +1000 mV Resolution 1 mV Accuracy ±5 mV ±0.5% Typical EMC ±6 mV ±6 mV Deviation ±0.2 mA ±0.2 mA Installation Cat. Input 4 to 20 mA Calibration Slope: 90 to 110% Slope trimmer Set Point Relay One, Isolated, 2 A, max 240 V, resistive load, 1.000.000 strokes...
  • Page 34 SPECIFICATIONS HI 931501 HI 931501 Range 0.00 to 14.00 pH Resolution 0.01 pH Accuracy ±0.02 pH Typical EMC ±0.1 pH Deviation ±0.2 mA Installation Cat. Input Offset: ±2 pH ∆O trimmer Calibration Slope: 80 to 110% Slope trimmer Temperature Fixed or automatic with PT100 Compensation from 0 to 100°C Readout...
  • Page 35: Initial Preparation

    INITIAL PREPARATION • Connect a 3-wire power cable to the 4-screw termi- FUSE nal strip; according to the FUSE voltage level as indicated and pay particular attention to the correct live, earth and neutral terminal connections. • For Model E, HI 931500, ELECTRODE HI 931501 and HI 931500 INPUT...
  • Page 36 • HI 8710 and HI 931500 are single dosage controllers for dosing either acid or alka- line liquid. If you plan to dose acid (e.g. in Hexavalent Chromium re- duction), make an open cir- cuit between ACID/BASE OPEN: ACID SHORT: BASE selection terminals (see rear panel descriptions #3 at page 12-13 and #4 at page 23).
  • Page 37 • Acid Contacts (HI 8711 and HI 931501 only): these two ACID contacts are for the connec- FUSE tion to the dosing pump for FUSE the acid. They act as a switch for the power to the drive. • Base contact (HI 8711 and HI 931501 only): these two BASE contacts are for the connec-...
  • Page 38: Operational Guide

    OPERATIONAL GUIDE The setting of the various keys are made via the front panel keys and trimmers. When each key is pressed the LED is lighted indi- cating to the user that the function is in operation. Make sure that the pH or the ORP meter with the electrode is calibrated before operat- ing the instruments (see page 40-42).
  • Page 39 Use a small screwdriver to adjust the trim- mers "ACID SET"/"COARSE" and "FINE" un- til the desired base set value is displayed. ACID BASE COARSE ACID SET FINE To set the working point of base dosage, press the "BASE SET" key. The display will show the set value for base dosage.
  • Page 40 Use a small screwdriver to adjust the trim- mers "∆AL" until the desired tolerance is displayed. ∆AL ∆AL For example: in HI 8710 if the set value is pH 3 and a ∆ Alarm of 1.5 pH is chosen, the instrument gives an alarm every time the measured pH value is higher than 4.5 pH or lower than 1.5 pH.
  • Page 41 When acid is dosed, the ACID LED will be lighted, and when DOSAGE SLOPE ACID BASE base is being dosed, the BASE ∆AL LED will be lighted (HI 8710, SURE COARSE HI 931500 and HI 931501 only). FINE When oxidants are dosed, the OXID LED will be lighted, and DOSAGE SLOPE...
  • Page 42: Ph Calibration

    SURE COARSE LED light is on) and not in the FINE set mode before proceeding the calibration (not for HI 8510). Note the temperature of the buffer with a ChecktempC or a glass thermometer. Remove the protective cap from the electrode, rinse it with some HI 7007 pH 7.01 solution (HI 7007), then...
  • Page 43 Note: to get accurate readings, use pH 4.01 if you are going to measure acid samples or pH 10.01 for alkaline mea- surements. Shake briefly and wait one minute before adjusting the slope trimmer to display pH4.01 (or 10.01) on the LCD if the temperature of the buffer solution is at 25°C, if not refer to the chart on page 43 for appropriate buffer value for the corresponding temperature.
  • Page 44: Orp Calibration

    ORP CALIBRATION Make sure you are in the mea- surement mode (MEASURE SURE COARSE LED light is on) and not in the FINE set mode before proceeding with the calibration (not for HI 8512). Immerse the electrode in HI 7020 ORP solution. HI 7020 Adjust the slope trimmer until an ORP value of between 200 mV and 250 mV is displayed.
  • Page 45: Ph Values At Various Temperature

    pH VALUES AT VARIOUS TEMPERATURE Temperature has an effect on pH. The calibration buffer solutions are effected by temperature changes to a lesser degree than normal solutions. Please refer to the following chart to perform the pH calibration: TEMP pH VALUES °C °F 4.01...
  • Page 46: Ph Diagnostic Tests

    DIAGNOSTIC TESTS The HI 8510, HI 8710 and HI 8711 are the only pH controllers with built-in autodiagnostic functions to enable the user to check and troubleshoot any malfunctions. The functions are made via front panel keys to isolate the cause of malfunction whether it is due to pH electrode contamination, inter- nal offset circuit or the amplifier circuit.
  • Page 47 B) Internal Offset Circuit Test Press the "pH7 TEST" key and the dis- play should show a value between 7 pH ±1 pH. This will verify the internal circuit of the meter in terms of the offset compen- sation. pH 7 TEST C) Amplifier Circuit Test Press the "pH4 TEST"...
  • Page 48: Orp Diagnostic Tests

    ORP DIAGNOSTIC TESTS The HI 8512 and HI 8720 are the only ORP controllers with built-in autodiagnostic func- tions to enable the user to check and trouble- shoot any malfunctions. The functions are made via front panel keys to isolate the cause of malfunction.
  • Page 49: Led Indication

    LED INDICATION The LEDs above all keys are designed to indicate the state of each function, whether it is active or the display is indicating the mode. For HI 8711 and HI 931501 only Each LED can be in one of the following states: A) Light stays on The mode is displayed...
  • Page 50: Taking Redox Measurements

    TAKING REDOX MEASUREMENTS Redox measurements allow the quantifica- tion of the oxidizing or reducing power of a solution, and are commonly expressed in Oxidation may be defined as the process during which a molecule (or an ion) loses electrons and reduction as the process by which electrons are gained.
  • Page 51 To make correct redox measurements the following conditions must prevail: The surface of the electrode must be cleaned and smooth. The surface of the electrode must un- dergo a preventive treatment depending on the solution to be measured has oxidizing or reductive characteristics. Because the Pt/PtO system depends on the pH, the pretreatment of the electrode may be determined by the pH and the redox potential...
  • Page 52 In the event that measurements are performed with solutions containing sulfides or proteins, the cleaning of the diaphragm of the refer- ence electrode must be performed (see page 20, "Cleaning Procedure"). In order to have a correct functioning of the ORP electrode, immerse it into HI 7020 and measure the response;...
  • Page 53: Electrode Conditioning And Maintenance

    ELECTRODE CONDITIONING AND MAINTENANCE Reference Filling Hole Reference Filling Hole Fill Hole Screw Sensitive Reference Wire Wire Sensitive Reference Wire Wire Reference Reference Junction Junction Glass Glass Bulb Bulb Plastic Body Glass Body pH Electrode pH Electrode Reference Reference Wire Wire Reference Reference...
  • Page 54 Replace the solution in the protective cap with a few drops of HI 70300 Storage Solu- tion or, in its absence, Filling Solution (HI7071 for single junction or HI7082 for double junction electrodes). AmpHel ® is a registered Trademark of "Hanna Instruments"...
  • Page 55 Follow the Preparation Procedure above be- fore taking measurements. Note: NEVER STORE THE ELECTRODE IN DISTILLED WATER OR DRY. PERIODIC MAINTENANCE Inspect the electrode and the cable. The cable used for the connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the elec- trode stem or bulb.
  • Page 56 IMPORTANT: After performing any of the cleaning procedures rinse the electrode thor- oughly with distilled water, drain and refill the reference chamber with fresh electrolyte, (not necessary for GEL filled electrodes) and soak the electrode in HI 70300 or HI 80300 Stor- age Solution for at least 1 hour before taking measurements.
  • Page 57 and then follow the Cleaning Procedure above. • For ORP Electrodes: polish the metal tip with a light abrasive paper (paying atten- tion not to scratch the surface) and wash thoroughly with water.
  • Page 58: Suggested Installations

    SUGGESTED INSTALLATIONS SHORT DISTANCE, INDOOR INSTALLATION Due to the low current involved, a very high grade of insulation is required. A dry environment is needed in order to ob- Ω. tain a level of insulation not lower than 10 MAXIMUM 10 METERS DRY ENVIRONMENT LESS THAN 80% RELATIVE HUMIDITY...
  • Page 59: Computer Interface

    MAXIMUM 300 METERS WET ENVIRONMENT RELATIVE HUMIDITY UP TO 100% pH-ORP METER pH METER WITH A pH TRANSMITTER AND CONVENTIONAL ELECTRODE MEASUREMENTS CAN BE TAKEN FROM DISTANCES UP TO 300 METERS (1000 FEET) AmpHel ® is a registered Trademark of "Hanna Instruments"...
  • Page 60: Accessories

    ACCESSORIES pH CALIBRATION SOLUTIONS HI 7004M pH 4.01 Buffer Solution, 230 mL HI 7004L pH 4.01 Buffer Solution, 460 mL HI 7006M pH 6.86 Buffer Solution, 230 mL HI 7006L pH 6.86 Buffer Solution, 460 mL HI 7007M pH 7.01 Buffer Solution, 230 mL HI 7007L pH 7.01 Buffer Solution, 460 mL HI 7009M pH 9.18 Buffer Solution, 230 mL HI 7009L pH 9.18 Buffer Solution, 460 mL...
  • Page 61 pH ELECTRODES HI 1090T Screwcap PG13.5 connector, double junction, glass-body PG13.5 THREAD φ 12mm φ 9.5mm 30mm 110mm HI 1110S Screw connector, single junction, glass-body HI 1130B/3 BNC connector, 3 m (9.9') cable, single junction, glass-body DIA 9.5mm 3/4 x 16 UNF M13 x 1.5 DIA 12mm DIA 20.5mm...
  • Page 62 HI 1210T Screwcap PG13.5 connector, double junction, plastic-body PG13.5 THREAD φ 12mm 30mm 110mm HI 1910B BNC connector, 1 m (3.3') cable, double junction, plastic-body, built-in amplifier 3/4 x 16 UNF DIA 12mm DIA 20.5mm 38.5mm 110mm HI 1911B BNC connector, 1 m (3.3') cable, double junction, plastic-body, built-in amplifier 3/4 x 16 UNF...
  • Page 63 HI 2910B/5 BNC connector, 5 m (16.5') cable, double junction, plastic-body, built-in amplifier 3/4 x 16 UNF DIA 12mm DIA 20.5mm 38.5mm 110mm ORP ELECTRODES HI 2930B/5 BNC connector, 5 m (16.5') cable, Pt, Ultem ® -body, built-in amplifier 3/4 x 16 UNF DIA 12mm DIA 20.5mm 38.5mm...
  • Page 64 HI 3210T Screwcap PG13.5 connector, Pt, plastic-body PG13.5 THREAD φ 12mm 30mm 110mm HI 3410S Screw connector, Pt, plastic-body HI 3430B/3 BNC connector, 3 m (9.9') cable, Pt, plastic-body 3/4 x 16 UNF DIA 12mm M13 x 1.5 DIA 20.5mm 38.5mm 110mm HI 3410S...
  • Page 65 OTHER ACCESSORIES ChecktempC Pocket-size thermometer with penetration probe and 0.1°C resolution (range -50.0 to 150.0°C) HI 76501/P Calibration Screwdriver (20 pcs) HI8614 pH Transmitter HI8614L pH Transmitter with LCD Display HI8615 ORP Transmitter HI8615L ORP Transmitter with LCD Display BL PUMPS Dosing Pumps with Flow Rate from 1.5 to 20 LPH HI 7871...
  • Page 66: Warranty

    WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The probes and the electrodes are warranted for a period of six months.
  • Page 67: Ce Declaration Of Conformity

    CE DECLARATION OF CONFORMITY DECLARATION OF CONFORMITY Hanna Instruments Srl V.le delle industrie 12 35010 Ronchi di Villafranca (PD) ITALY herewith certify that the process controllers HI 8510E HI 8510T HI 8512E HI 8512T HI 8710E HI 8710T HI 8720E...
  • Page 68 PRINTED IN MANPROCR1 PORTUGAL 07/96...

This manual is also suitable for:

Hi931501Hi 8720Hi 931500Hi 932500

Table of Contents