Setup Mode; Control Mode - Hanna Instruments HI 710 Series Instruction Manual

Process, panel-mounted, microprocessor-based, conductivity and tds controllers
Hide thumbs Also See for HI 710 Series:
Table of Contents

Advertisement

SETUP MODE

HI 700 and HI 710 offer a multitude of possibilities from
ON/OFF or PID dosage to analog recorder output and from
alarm to selftest features.
The Setup Mode allows the user to set all needed character-
istics of the meter.
The setup mode is entered by pressing SETUP
and entering the password when the device is
in idle or control mode.
Generally speaking, if the password is not inserted the user
can only view the setup parameters (except for password)
without modifying them (and the device remains in control
mode). An exception is certain setup items, or flags, which
can activate special tasks when set and confirmed.
Each setup parameter (or setup item) is assigned a two-digit
setup code which is entered and displayed on the secondary
LCD.
The setup codes can be selected after password and CFM
are pressed. When CFM is pressed, the current setup item is
saved on EEPROM and the following item is displayed.
Whenever LCD is pressed, the de-
vice reverts back to control mode.
The same is true when CFM is
pressed on the last setup item.
The possible transitions in setup mode are the following:
ENTERING THE PASSWORD
• Press SETUP to enter the setup mode. The LCD will display
"0000" on the upper part and "PAS" on the lower. The first
digit of the upper part of the LCD will blink.
• Enter the first value of the pass-
word by the
12
or
keys.
• Then confirm the displayed digit with
and move to the next one.
• When the whole password has been
inserted, press CFM to confirm it.
Note
The default password is set at "0000".
• The LCD will display "SET" on the
upper part and "c.00" on the lower,
allowing the user to edit setup param-
eters (see table below).
• Enter the code of the parameter you want to set, using the
arrow keys as per the password procedure above (e.g. 41).
• Confirm the code by pressing CFM and the default or the
previously memorized value will be displayed with the first
digit blinking.
Note
When the password is not inserted or a wrong password is
confirmed, the display will only show the previously memo-
rized value, without blinking (read only mode). In this case,
the value cannot be set. Press LCD and start again.
• Enter the desired value using the arrow keys and then press
CFM.
13