EC VALUES AT VARIOUS TEMPERATURES

Temperature has a significant effect on conductivity. Table
below shows EC values at various temperatures for the Hanna
calibration solutions.
TEMPERATURE
°C
°F
0
32
5
41
10
50
15
59
16 60.8
17 62.6
18 64.4
19 66.2
20
68
21 69.8
22 71.6
23 73.4
24 75.2
25
77
26 78.8
27 80.6
28 82.4
29 84.2
30
86
31 87.8
50
EC VALUES (μS/cm)
HI7030
HI7031 HI7033 HI7034 HI7035
HI8030
HI8031 HI8033 HI8034 HI8035
7150
776
64
48300
65400
8220
896
65
53500
74100
9330
1020
67
59600
83200
10480
1147
68
65400
92500
10720
1173
70
67200
94400
10950
1199
71
68500
96300
11190
1225
73
69800
98200
11430
1251
74
71300
100200
11670
1278
76
72400
102100
11910
1305
78
74000
104000
12150
1332
79
75200
105900
12390
1359
81
76500
107900
12640
1386
82
78300
109800
12880
1413
84
80000
111800
13130
1440
86
81300
113800
13370
1467
87
83000
115700
13620
1494
89
84900
117700
13870
1521
90
86300
119700
14120
1548
92
88200
121800
14370
1575
94
90000
123900

PROBE MAINTENANCE

PERIODIC MAINTENANCE
HI7039
HI8039
2760
3180
CLEANING PROCEDURE
3615
4063
4155
4245
4337
4429
4523
4617
4711
4805
4902
5000
5096
5190
5286
5383
5479
5575
Probe can be compensated for normal contamination by a
process of recalibration. When calibration can no longer be
achieved, remove the conductivity probe from the system for
maintenance.
Inspect the probe and the cable. The cable used for the con-
nection to the controller must be intact and there must be no
points of broken insulation on the cable.
Connectors must be perfectly clean and dry.
Rinse the probe with tap water. If a more thorough cleaning
is desired, remove the sleeve and clean the platinum sensors
with a non-abrasive cloth or HI7061 cleaning solution. Re-
insert the sleeve in the same direction as before.
Recalibrate the instrument before reinserting the probe in the
system.
Note
Always recalibrate the instrument when attaching a new probe.
51