4.8 Meter (DMM) Tests

4.8.1 Meter DC Voltage Accuracy Test

Proceed as follows to test the meter dc voltage measurement accuracy:
1. Connect the test tool to the 5500A as for the previous test (see Figure 4-18).
2. Select the following test tool setup:
3. Set the range to the first test point in Table 4-12.
4. Set the 5500A to source the appropriate dc voltage.
5. Observe the reading and check to see if it is within the range shown under the
6. Continue through the test points.
7. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
4-28
Dangerous voltages will be present on the calibration source
and connecting cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
Press
(this key will toggle the menu bar on and off if the test tool is
already in the meter mode)
Press
to open the Measurement menu, and select
Press
to select MANUAL ranging; use
appropriate column.
Table 4-12. Meter Volts dc Measurement Verification Points
Range
500.0 mV
5.000 V
50.00 V
500.0 V
1100 V
WARNING
5500A output V dc
+ 500 mV
- 500 mV
0 mV
+ 5.000 V
- 5.000 V
+ 50.00 V
- 50.00 V
+ 500.0 V
- 500.0 V
+ 1000 V
- 1000 V
V dc
to select the ranges.
Meter Reading
497.0 to 503.0
-497.0 to -503.0
-0.5 to +0.5
4.970 to 5.030
-4.970 to -5.030
49.70 to 50.30
-49.70 to -50.30
497.0 to 503.0
-497.0 to -503.0
0.990 to 1.010
-0.990 to -1.010