Input B Lf-Hf Gain Calibration Points - Fluke 199B Service Manual

Scopemeter
Hide thumbs Also See for 199B:
Table of Contents

Advertisement

Fluke 19xB-19xC-2x5C
Service Manual
3. Set the 5500A SCOPE output to source the signal required for the first calibration
4. Set the 5500A in operate (OPR) or standby (STBY) as indicated.
5. Press
6. Wait until the display shows calibration status
7. Press
8. When you are finished, set the 5500A to Standby.
9. Continue at Section 5.6.3.
Cal step
CL 0674
CL 0410
CL 0724
CL 0421
CL 0482
CL 0483
5-10
Figure 5-4. 5500A SCOPE Output to Test Tool Input B
point in Table 5-2.
to start the calibration.
to select the next calibration step, set the 5500A to the next calibration
point signal, and start the calibration.
Continue through all calibration points of Table 5-2.
Table 5-2. Input B LF-HF Gain Calibration Points
UUT input signal
none
0.5 Vpp square wave, 1 kHz
none
0.5 Vpp square wave, 1 kHz
0.5 Vpp sine wave, 50 kHz
0.5 Vpp sine wave
Fluke 199B-C, 225C: 221 MHz
Fluke 196B-C, 215C: 141 MHz
Fluke 192B-C:
:READY
STANDBY
SCOPE edge, 0.5 Vpp, 1 kHz
STANDBY
SCOPE edge, 0.5 Vpp, 1 kHz
SCOPE levsine, 0.5 Vpp, 50 kHz
SCOPE levsine, 0.5 Vpp,
91 MHz
.
5500A Setting
221 MHz
141 MHz
91 MHz
al55scb.bmp