Input A&B Position; Input A&B Gain Calibration Points - Fluke 199B Service Manual

Scopemeter
Hide thumbs Also See for 199B:
Table of Contents

Advertisement

Fluke 19xB-19xC-2x5C
Service Manual
Cal step
CL 0660
CL 0604
CL 0637
CL 0504
CL 0624
CL 0599
Not for software versions
V05.01 and V05.02
CL 0600
CL 0601
CL 0602
CL 0603
CL0662
CL 0605
CL 0606
CL 0607
CL0664
CL 0608
CL 0609
CL 0610
5.6.4 Input A&B Position
Proceed as follows to do the Input A&B Position calibration:
1. Press
2. Remove all Input A and Input B connections (open inputs).
3. Press
4. Wait until the display shows calibration status
5. Continue at Section 5.6.5
5-12
Table 5-3. Input A&B Gain Calibration Points
UUT input value (5500A SCOPE, MODE volt, SCOPE Z 1 MΩ, 1 kHz)
300 mV
500 mV
500 mV
10 mV
25 mV
50 mV
100 mV
250 mV
2.5 V
to select calibration adjustment step
and V05.02), or CL 0619 (software versions V05.03 and newer).
to start the calibration
none
(5500 standby)
none
(5500 standby)
2 V
1 V
5 V
20 V
10 V
25 V
50 V
(set 5500A to OPR!)
CL 0620 (software versions V05.01
:READY.