Resistance Measurement Verification Points - Fluke 199B Service Manual

Scopemeter
Hide thumbs Also See for 199B:
Table of Contents

Advertisement

Figure 4-19. Test Meter Tool Input to 5500A Normal Output 4-Wire
2. Select the following test tool setup:
Press
Press
to open the Measurement menu, and select
Press
to select AUTO ranging.
3. Set the 5500A to source the appropriate resistance value for the first test point in
Table 4-14.
Use the 5500A "COMP 2 wire" mode for the verifications up to and including
50 kΩ. For the higher values, the 5500A will turn off the "COMP 2 wire" mode.
4. Observe the reading and check to see if it is within the range shown under the
appropriate column.
5. Continue through the test points.
6. When you are finished, set the 5500A to Standby.
Table 4-14. Resistance Measurement Verification Points
5500A output
400Ω
4 kΩ
40 kΩ
400 kΩ
4 MΩ
30 MΩ
Performance Verification
4.8 Meter (DMM) Tests
Ohms
Meter Reading
0.0 to 0.5
397.1 to 402.9
3.971 to 4.029
39.71 to 40.29
397.1 to 402.9
3.971 to 4.029
29.77 to 30.23
4
al55ex4w.bmp
4-31