5.2 Calibration Procedure Steps

To do a complete calibration adjustment you must do all following steps:
1. Select the Calibration Mode, section 5.3
2. Do the Contrast Calibration Adjustment, section 5.4
3. Do the WarmingUp & PreCalibration, section 5.5
4. Do the Final Calibration, section 5.6
5. Save the Calibration Data and Exit the calibration mode, section 5.7
6. Do the probe Calibration, section 5.8
Steps 4 and 5 must always be done together and completely in the order given in this
manual. Skipping adjustment steps will make the calibration invalid!
The following partial calibrations are allowed:
• Contrast calibration, do the above-mentioned steps 1, 2, and 5.
• Probe calibration, do the above-mentioned step 6.

5.3 Starting the Calibration

Follow the steps below to start the calibration:
1. Power the test tool via the power adapter input using the BC190 power adapter.
2. Check the actual test tool date, and adjust the date if necessary (the calibration date
3. Select the calibration mode.
5-4
Fluke 5500A Multi Product Calibrator, including SC300 or SC600 Oscilloscope
Calibration Option.
Stackable Test Leads (4x), supplied with the 5500A.
50Ω Coax Cable (2x), for example Fluke PM9091 (1.5m) or PM9092 (0.5m).
50Ω feed through termination, Fluke PM9585.
Male BNC to Dual Female BNC Adapter (1x), Fluke PM9093/001.
Dual Banana Plug to Female BNC Adapter (1x), Fluke PM9081/001.
If during normal operation the display cannot be made dark or light enough, or if the
display after a test tool reset is too light or too dark, you can do this calibration.
The probe calibration matches the probe to the used input channel.
will become the test tool date when saving the calibration data):
Press
(toggles the menu bar on-off)
press
to open the
using
select
press
to open the
adjust the date if necessary.
The Calibration Adjustment Procedure uses built-in calibration setups, that can be
accessed in the calibration mode.
menu
OPTIONS
...
DATE ADJUST
menu
DATE ADJUST