Fluke 19xB-19xC-2x5C
Service Manual
4.9 Probe Calibration Generator Test
To verify the internal probe calibration square wave generator, you can do a Probe
Calibration as described in section 5.8. If no square wave appears on the screen, either
• the probe is defective: try another probe, check the probe with an external voltage in a
or
• the internal square wave generator is defective.
This is the end of the Performance Verification Procedure.
4-32
scope application,