Frequency Sourcing - Fluke 744 User Manual

Documenting process calibrator
Hide thumbs Also See for 744:
Table of Contents

Advertisement

744
Users Manual

Frequency Sourcing

Range
0.00 Hz to 10.99 Hz
11.00 Hz to 109.99 Hz
110.0 Hz to 1099.9 Hz
1.100 kHz to 21.999 kHz
22.000 kHz to 50.000 kHz
Waveform Choices: Zero-symmetric sine or positive square wave, 50% duty cycle.
Amplitude: 0.1 to 10 V pk
Amplitude Accuracy:
0 Hz to 1099 Hz: 3% of output + 0.5% f.s
1.1 kHz to 10.9 kHz: 10% of output + 0.5% f.s
11 kHz to 50 kHz: 30% of output + 0.5% f.s
Maximum Input Voltage: 30 V dc
112
Accuracy
1 and 2 Year
0.01 Hz
0.01 Hz
0.01 Hz
0.002 kHz
0.005 kHz