Fluke 744 User Manual page 125

Documenting process calibrator
Hide thumbs Also See for 744:
Table of Contents

Advertisement

Temperature, Thermocouples (cont)
Type
Range C
U
-200 to 0
0 to 600
Sensor inaccuracies not included.
Accuracy with external cold junction; for internal junction add 0.2 C
Resolution: 0.1 C
Temperature Scale: ITS-90 or IPTS-68, selectable
Compensation: ITS-90 per NIST Monograph 175 for B,R,S,E,J,K,N,T; IPTS-68 per IEC 584-1 for B,R,S,E,J,K,T;
IPTS-68 per DIN 43710 for L,U.
Temperature Coefficient: 0.05 C/ C in the range -10 to 18 C and 28 to 50 C
Common Mode Error: 0.01 C/(Common Mode Volt)
Maximum Input Voltage: 30 V
Measure °C
1 Year
2 Year
0.6
0.9
0.3
0.4
Documenting Process Calibrator
Specifications
Source °C
1 Year
2 Year
0.4
0.6
0.3
0.4
115

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents