Instrument I/D And Setup - Fluke 8508A User Manual

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Users Manual
Power On and Reset Conditions
Cleared (i.e., nothing enabled).
Clear Status
*CLS
Clears all the event registers and queues except the output queue. The output queue and
MAV bit will be cleared if *CLS immediately follows a '"Program Message
Terminator".
Power On Status Clear
*PSC
Sets the flag controlling the clearing of defined registers at Power On.
Nrf
A number which, when rounded to an integer value of zero, sets the power on clear flag
false. This allows the instrument to assert SRQ at power on.
When the value rounds to an integer value other than zero it sets the power on clear flag
true, which clears the standard event status enable and service request enable registers so
that the instrument will not assert an SRQ on power up. This data is saved in non volatile
memory for use at Power On.
Recall Status Clear Flag
*PSC?
(Nr1) Recalls the saved Power On status condition: false (0) and true (1).

Instrument I/D and Setup

I/D (Identification)
IDN?
Recalls the instrument's manufacturer, model number, serial number and firmware level.
4-44
*
CLS
Clear Status
*
PSC
phs
Power On Status Clear
*
PSC?
Recall Power On Status Clear)
*
IDN?
I/D
Nrf
adj265f.eps
adj264f.eps
adj267f.eps
adj269f.eps

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents