Execute Trigger; Execute Trigger And Take A Reading; Settling Delay - Fluke 8508A User Manual

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Users Manual

Execute Trigger

This command conforms to the IEEE 488.2 standard requirements.
*TRG
Equivalent to a Group Execute Trigger (GET). Causes a single reading to be taken.

Execute Trigger and Take a Reading

X?
Equivalent to performing an Execute Trigger (
X? is intended for high speed use.

Settling Delay

Nrf
A numeric value for the required settle delay. The minimum period allowed is 0, and the
maximum is 65,000 seconds.
Examples:
The programmed delay is active with TRG_SRCE EXT selected, although delays may be
programmed whilst unit is in Remote with default (internal) triggers selected. They will
then become active upon the selection of the external trigger.
Execution Errors:
Execution errors are generated if an attempt is made to program the delays when the
instrument is not in remote control.
An execution error is generated if the selected value of Nrf exceeds the limiting value.
Power On and Reset Conditions
DFLT
4-34
Execute Trigger and Take Reading
DELA Y
sets a settle delay after trigger of 1ms before the reading begins
DELAY 0.001
sets the default delay for the selected function, range, filter etc.
DELAY DFLT
Delay Selection Resolution
<0.01 s
10 µs
0.01 s to 0.1 s
100 µs
0.1 s to 1 s
1 ms
1 s to 10 s
10 ms
>10 s
100 ms
TRG

Execute Trigger

X?
*TRG;RDG?
phs
Nrf
DFLT
Setting Delay
) followed by a reading query.
adj245f.eps
adj246f.eps
adj247f.eps