Fluke 8508A User Manual page 124

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Users Manual
"Special" measurement commands for DCV Linearity Calculations
HV_LIN?
Triggers a measurement and records the result for use with the special calibration
command CHSE? HV_LIN. The signal supplied for each measurement should match the
supplied parameter.
SRC_N1K
Measurement source is -1000V DC.
SRC_N500
Measurement source is -500V DC.
ZERO
Measurement source is 0.0V DC.
SRC_P500
Measurement source is +500V DC.
SRC_P1K
Measurement source is +1000V DC.
Response
(Nr1) The value returned identifies the success (0) or failure (1) of the measurement.
Execution Errors
4-50
HV_LIN?
"Special" measurement commands for DCV Linearity Calculations
If special calibration is not enabled.
S R C _ N 1 K
phs
SRC_N500
ZERO
SRC_P500
SRC_P1K
adj361f.eps