Fluke 8508A User Manual page 68

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Users Manual
Right Di
splay
Line
SpotF
Ser#
Due
Dp Style
IEEE 488 ADDRESS
The menu displays the present address value, and the numeric-keyboard keys are
activated. Any valid numeric value (0-30) may be entered.
LINE Frequency
This selection adjusts the integration time of the Analogue to Digital convertor to align
with an integral number of power line cycles at 50 Hz or 60 Hz. Correct setting is critical
to the Normal Mode and Common Mode rejection of the instrument. Once set, the
frequency is retained even after the Multimeter is turned OFF.
SPOTF Menu
This display is for information only. The Spot Frequencies cannot be altered except via
Calibration at new values.
The
of the six possible spot frequencies that could have been calibrated for the currently
active ACV range.
3-34
Provides a contrast menu for adjustment to the right hand display.
Provides selection of the power line frequency for the Multimeter.
When in ACV function this softkey displays the
spot frequencies at which the Multimeter has been calibrated.
Displays the
SER
version of the software in it.
Displays the
CAL DUE
for recalibration of the instrument.
Provides selection for display styles for the measurements.
IEEE Bus address key Menu
menu has six soft keys (
SPOTF
# menu, to view the serial number of the Multimeter and
date. This shows the user-entered recommended date
Line Frequency
SPOTF Menu
through
Sp1
Sp6
menu to view the
SPOTF
), each allows the user to view one
adj120f.eps
adj121f.eps
LINE
adj122f.eps

Hide quick links:

Advertisement

Table of Contents
loading