Zero Point
1. On the 8508A, select the required range.
2. On the Calibrator, select zero
3. On the 8508A, press
4. Press
Range Gain Point
1. On the Calibrator, select Range Gain point Output value. Alternatively if using a
2. On the 8508A, Press
3. Select
4. Use the numeric keys with the
5. Press
6. On the Calibrator, set
7. Press the
HiV Mode Calibration
The Multimeter can deliver a lethal electric shock when operating in the
High Voltage Resistance mode. Voltages up to 240V may appear on the
terminals.
To avoid electric shock when performing the following calibration
procedures:
1. Press the
2. Press the
Range Zero and Gain Calibration Procedure
After the initial setup and connecting up, use the following procedure to calibrate range
zero and range gain for the ranges and points detailed in Table 6-5 for the HiV mode.
Just one range can be calibrated if required, but for a full calibration start with the 20 M
range and work up to the 20 G range.
6-22
resistor connect the standard resistor for a 4-wire resistance zero.
CAL
to initiate the calibration operation.
SAMPLE
Calibration is complete when the Busy legend goes out and the calibrated
measurement is displayed.
standard resistor re-connect the standard resistor to measure its resistance.
CAL
on the
Set
CALIBRATION menu
the calibrator (or the value of the resistance standard) at the range gain value, then
press Enter.
to initiate the calibration operation.
SAMPLE
Calibration is complete when the Busy legend goes out and the calibrated
measurement is displayed.
STANDBY
key to revert to the ranges menu.
Never touch any lead or terminal unless you are absolutely certain
that no dangerous voltage is present.
Make sure that signal leads are in a safe condition before you
handle them in any way.
key and select
Plus
key and select
CONFIG
and Operate. Alternatively if using a standard
to enter the
CALIBRATION
to revert to
CALIBRATION
.
menu to key in the true output value of
SET VALUE
.
XW Warning
From the
HiV .
HiV
and
.
Filt
4w
menu.
menu.
menu select the
20 M
range.