INT / EXT
This selects the connection of the multimeter's guarding. (For backwards compatibility
LCL & REM can also be used.)
For scan operations, the guard selection is applied to the channel currently being applied
to the A-D converter.
Power On and Reset Conditions
INT (internal)

Measurement Gate Width

This command selects the gate width for frequency readings during AC measurements.
FAST_ON
Selects a gate width of 50ms, and a frequency resolution of 4.5 digits.
FAST_OFF
Selects a gate width of 1s, and a frequency resolution of 6.5 digits.
Both selections are mutually exclusive.
The use of the longer gate width results in a 6.5 digit frequency measurement. The
frequency gate is triggered at the same point as the A-D conversion which could be
significantly shorter than 1 second. This may reduce the read-rate, as the measurement
processing cannot begin until both the frequency gate and the A-D conversion are
complete.
Power On and Reset Conditions
FAST_ON.

Maximum, Minimum and Peak-Peak

Recall Stored Values
MAX?
Recalls the stored value of the maximum signal value to be measured since the most-
recent general reset, store reset or function change.
MIN?
Recalls the stored value representing the minimum signal value to be measured since the
most-recent general reset, store reset or function change.
4-22
phs
GA TE

Measurement Gate Width

MA X?
MIN?
PKPK?
Recall Stored values
FA ST_ON
FA ST_OFF
adj220f.eps
adj221f.eps