Fluke 8508A User Manual page 100

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Users Manual
Enable Averaging
All selections are mutually exclusive.
AVG AV...
Averages the number of readings requested (4, 16, or 64) as a rolling average.
From a cleared average memory, the average is the mean of the number of
readings to date, until the selection window number is reached. The average
memories are cleared on each command update.
AVG BLOC_N
Selects N readings for a block average.
The parameter BLOC_N selects the average of N readings, where only one
result is obtained after the required number of readings (internally
triggered) have been obtained.
AVG OFF
Turn averaging OFF; the number N is not destroyed.
Power On and Reset Conditions
OFF
Set Block Size
Sets the integer constant N for use with the averaging math function.
Nrf
An integer value to be used as the number of readings to be taken and averaged in each
block.. This value is saved in non volatile memory.
Example:
N 15
Execution Errors
Execution errors will be generated when N > 10,000.
Power On and Reset Conditions: No Change. The number N is saved at Power Off.
4-26
AVG
N
sets the number of readings to be averaged to 15.
phs
AV4
AV16
AV64
BLOC_N
OFF
Enable Averaging
Note
Note
phs
Nrf
Set Block Size
adj226f.eps
adj227f.eps