8508A
Users Manual
DC Current
Press the
DCI
Auto
DCI Configuration (Resolution, LP Filtering and Fast)
Press the
are toggle keys.
Resl
Filt
Fast
DCI Resolution
Press the
5
6
7
Refer to table 3-1 for the ADC integration times applicable to the various resolution and
Fast On/Off settings.
To revert to the
3-20
key to show the
DCI
This function provides measurements, in decade ranges from 200 A to
20A. (NOTE: the 20A range is only available on the front terminals. The
front current terminal is not fused.)
To cancel autoranging and enter a manually selected range, press any range
selection or the Auto softkey. The Multimeter enters the selected range or
reverts to the auto-selected range.
key to see the
Config
Provides selection of the resolution for subsequent readings.
Controls a two-pole analog filter for increased noise rejection.
The Filter annunciator on the left-hand display indicates that the filter is
active and indicates its cut-in frequency.
Select a higher read rate. This will result in an increase in uncertainty due
to noise.
The Fast annunciator on the left-hand display indicates that the higher read
rate is active.
key to see the
Resl
5-1/2-digits resolution
6-1/2-digits resolution
7-1/2-digits resolution
DCI CONFIG
menu.
DCI
DCI Menu
menu.
DCI CONFIG
Resl
DCI Configuration Menu
menu:
DCI RESL
DCI RESL
Menu
menu, press the
Config
is a menu keys;
Filts and Fast
key.
adj097f.eps
adj098f.eps
adj099f.eps