Fluke 8508A User Manual page 117

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

ESE?
(Nr1) Recalls the enable mask for the standard defined event register. The returned
value, when converted to base 2 (binary), identifies the bits as defined in the IEEE 488.2
standard and detailed in the status data structure diagram.
Power On and Reset Conditions
The Power ON condition depends on the condition stored by the common *PSC
command - if 0 then it is not cleared; if 1 then the register is cleared. Reset has no effect.
Read Measurement Event Register
MESR?
(Nr1) Destructively reads the measurement event register which, when converted to base
2 (binary), identifies the events that have occurred since the most-recent read or general
clear of this register. The detail is contained in the status data structure diagram. The
register is also cleared by the common command CLS.
Power On and Reset Conditions
The register is cleared.
Measurement Event Enable
MESE
Enables the measurement event bits which will generate a summary message in the
service request byte. See the device status reporting model for detail.
Nrf
A number which, when rounded to an integer and expressed in base 2 (binary), enables
the appropriate bits in this event enable register.
Recall Measurement Event Enable
This measurement event status data structure conforms to the IEEE 488.2 standard
requirements for this structure.
MESE?
(Nr1) Recalls the measurement status register enable mask which, when converted to
base 2 (binary), identifies the enabled bits which will generate a summary message in the
service request byte.
Remote Operations Using the IEEE 488 Interface
MESR?
Read Measurement Event Register
MESE
phs
Measurement Event Enable
MESE?
Recall Measurement Event Enable
Nrf
4
Status Reporting
adj258f.eps
adj257f.eps
adj256f.eps
4-43