Procedure
Repeat the following sequence for each range, starting with the 200 A range and
working up to the 20A range, as listed in Table 6-12.
1. Select the required 8508A range.
2. On the calibrator set the output to the current and frequency for the point being
verified as listed in Table 6-12. Set the output to
3. Note the 8508A reading. Compare the result with the allowable tolerance, taking into
account any known error from nominal of the calibrator for this output.
4. Repeat steps 2 – 4 for each point to be verified on the 8508A range to be verified.
5. Repeat steps 1 – 5 for each 8508A range to be verified.
6. Set the calibrator output to
Table 6-12. AC Current Verification Points and Sequence
8508A Range
200 A
2 mA
20 mA
200 mA
2 A
20 A
1. Based on 365 day Tcal 1 C 99 %CL Absolute specifications. See comments in Applicable Verification Tolerance section above
regarding the impact of calibration uncertainties and use of other specifications.
.
STANDBY
Current
Frequency
300 Hz
100 A
1 kHz
100 A
3 kHz
100 A
5 kHz
100 A
10 kHz
100 A
1 mA
300 Hz
1 mA
1 kHz
1 mA
3 kHz
1 mA
5 kHz
1 mA
10 kHz
10 mA
300 Hz
10 mA
1 kHz
10 mA
3 kHz
10 mA
5 kHz
10 mA
10 kHz
100 mA
300 Hz
100 mA
1 kHz
100 mA
3 kHz
100 mA
5 kHz
100 mA
10 kHz
1 A
300 Hz
1 A
1 kHz
1 A
3 kHz
1 A
5 kHz
1 A
10 kHz
10 A
300 Hz
10 A
1 kHz
10 A
3 kHz
10 A
5 kHz
10 A
10 kHz
Calibration and Verification
Performance Verification
.
OPERATE
Tolerance
0.0830 A
0.0830 A
0.0830 A
0.0830 A
0.0830 A
0.000580 mA
0.000580 mA
0.000580 mA
0.000580 mA
0.000580 mA
0.00580 mA
0.00580 mA
0.00580 mA
0.00580 mA
0.00580 mA
0.0545 mA
0.0545 mA
0.0545 mA
0.0545 mA
0.0545 mA
0.000945 A
0.000945 A
0.001055 A
0.001055 A
0.001055 A
0.01140 A
0.01140 A
0.02740 A
0.02740 A
0.02740 A
6
1
6-41