Dc Voltage - Fluke 8508A User Manual

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

DC Voltage

Range
Full Scale
95 % Confidence Level
200 mV
199.990 000
2 V
1.999 900 00
20 V
19.999 000 0
200 V
199.990 000
1000 V
1050.000 00
99 % Confidence Level
200 mV
199.990 000
2 V
1.999 900 00
20 V
19.999 000 0
200 V
199.990 000
1000 V
1050.000 00
Range
200 mV
2 V
20 V
200 V
1000 V
Type ....................................................................... Multi-slope, multi-cycle A-D Converter
[5]
CMRR (1 k
unbalance)
.................................... 140 dB at DC and 1 - 60 Hz
[5]
NMRR
Filter Out ................................................................. 60 dB at 50/60 Hz ±0.09 %
Filter In .................................................................... 110 dB at 50/60 Hz ±0.09 %
Protection (All ranges)........................................... 1 kV rms
Input Impedance
200 mV to 20 V Ranges.......................................... > 10 G
200 V & 1000 V Ranges ......................................... 10.1 M ± 1 %
Max Input Current ................................................. 50 pA
Ratio Accuracy
Range to Range...................................................... ±(Net Front Input Accuracy + Net Rear Input Accuracy)
Within Range .......................................................... Apply 24 hour or 20 minute Transfer Uncertainty specifications
Settling Time (to 10 ppm step size)
Filter Out ................................................................. < 50 ms
Filter In .................................................................... < 1 s

DC Voltage

Uncertainty Relative to Cal Stds
[15]
24 hour
90 day
TCal ±1 °C
TCal ±1 °C
0.7 + 0.5
1.4 + 0.5
0.5 + 0.2
1.4 + 0.2
0.5 + 0.2
1.4 + 0.2
1.0 + 0.2
2.6 + 0.2
1.0 + 0.5
2.6 + 0.5
0.8 + 0.6
2.0 + 0.6
0.6 + 0.25
1.8 + 0.25
0.6 + 0.25
1.8 + 0.25
1.2 + 0.25
3.5 + 0.25
1.2 + 0.6
3.5 + 0.6
DC Voltage (Secondary Specifications)
Transfer Uncertainty
20 mins ±1 °C ± (ppm
Reading+ ppm Range)
0.4 + 0.3
0.12 + 0.1
0.12 + 0.1
0.4 + 0.1
0.4 + 0.3
[1] [2] [3]
± (ppm Reading + ppm Range)
365 day
TCal ±1 °C
TCal ±1 °C
2.7 + 0.5
4.5 + 0.5
2.7 + 0.2
3.0 + 0.2
2.7 + 0.2
3.0 + 0.2
4.0 + 0.2
4.5 + 0.2
4.0 + 0.5
4.5 + 0.5
3.5 + 0.6
6.0 + 0.6
3.5 + 0.25
4.0 + 0.25
3.5 + 0.25
4.0 + 0.25
5.2 + 0.25
6.0 + 0.25
5.2 + 0.6
6.0 + 0.6
[1] [2] [3]
Temperature Coefficient
15 °C - 30 °C
± ppm Reading/°C
0.4
0.3
0.3
0.7
0.7
Specifications
Specifications
Absolute Uncertainties
[4]
365 day
365 day TCal
±5 °C
5.0 + 0.5
3.5 + 0.2
3.5 + 0.2
5.5 + 0.2
5.5 + 0.5
6.5 + 0.6
4.5 + 0.25
4.5 + 0.25
7.0 + 0.25
7.0 + 0.6
5 °C - 15 °C
30 °C - 40 °C
0.6
0.5
0.5
1.0
1.0
5
5-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents