Table of Contents

Advertisement

Nrf
A numeric value that selects the range for the expected signal. For example, an
10, or even 15.6789, selects the 20 V range. Any valid numeric value cancels autorange.
AUTO
Selects the autorange mode where the multimeter will determine the appropriate range for
the measured signal.
If the signal exceeds the maximum capability then
front panel, and the appropriate bit is set in the device status registers.
FILT_ON / FILT_OFF
inserts a hardware analog filter into the signal path.
FILT_ON
RESL5 / RESL6 / RESL7 / RESL8
Sets the resolution of the measurement to 5.5, 6.5, 7.5 or 8.5 digits.
FAST_ON / FAST_OFF
Reduces the A-D integration time, for faster conversions.
FAST_ON
TWO_WR
Requires only a connection to the input terminals.
FOUR_WR
Requires a connection to the input terminals and to the sense terminals.
Power On and Reset Conditions
1kV, FILT_OFF, RESL7, FAST_ON, TWO_WR

AC Voltage

The following commands are used to select the
Remote Operations Using the IEEE 488 Interface
ACV
phs
Syntax Diagrams of Major 8508A Functions
RANGE OVERLOAD
function and set its configuration.
ACV
,
Nrf
AUTO
FILT100HZ
FILT40HZ
FILT10HZ
FILT1HZ
RESL5
RESL6
TFER_ON
TFER_OFF
DCCP
ACCP
SPOT_ON
SPOT_OFF
TWO_WR
FOUR_WR
of 2,
Nrf
appears on the
adj210f.eps
4
4-11