Step
Test Function
11
Press
µA mode
12
Turn the Rotary Switch to the
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect the calibrator to hanheld multimeter's A and COM terminals before applying 3A and 10A
Move the test lead from A and COM terminal to uA.mA and COM terminal
13
Press
mA mode
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect the calibrator to hanheld multimeter's A and COM terminals before applying 3A and 10A
14
Press
A mode
Agilent U1251B/U1252B User's and Service Guide
Cal Range
button to go to
500 µA
5000 µA
Open
position
50 mA
440 mA
5 A
10 A
button to go to
50 mA
440 mA
button to go to
5 A
10 A
Performance Tests and Calibration
Input
30 µA,1 kHz
030.00 µA
300 µA,1 kHz
300.00 uA
300 µA,1 kHz
0300.0 µA
3000 µA, 1 kHz
3000.0 µA
Input terminal
oPEn
open (remove any
test leads and
Short Plugs from
the input terminal)
30 mA
30.000 mA
300 mA
300.00 mA
3 A
3.0000 A
10 A
10.000 A
3 mA, 1 kHz
03.000 mA
30 mA, 1 kHz
30.000 mA
30 mA, 1 kHz
030.00 mA
300 mA, 1 kHz
300.00 mA
0.3 A, 1 kHz
0.3000 A
3 A, 1 kHz
3.0000 A
3 A, 1 kHz
3.0000 A
10 A, 1 kHz
10.000 A
Cal Item
U1251B
U1252B
030.00 µA
300.00 µA
0300.0 µA
3000.0 µA
oPEn
30.000 mA
300.00 mA
3.0000 A
10.000 A
03.000 mA
30.000 mA
030.00 mA
300.00 mA
0.3000 A
3.0000 A
3.0000 A
10.000 A