Step
Test Function
7
Press
the
mV mode
Agilent U1251B/U1252B User's and Service Guide
Range
button to go to
50 mV
[1]
500 mV
1000 mV
Performance Tests and Calibration
5520A Output
U1251B
50 mV, 1 kHz
± 0.34 mV
50 mV, 10 kHz
± 0.86 mV
45 mV, 20 kHz
N/A
50 mV, 30 kHz
± 0.86 mV
45 mV, 100 kHz
N/A
500 mV, 45 Hz
± 3.25 mV
500 mV, 1 kHz
± 3.25 mV
500 mV, 10 kHz
± 8.6 mV
450 mV, 20 kHz
N/A
500 mV, 30 kHz
± 8.6 mV
450 mV, 100 kHz
N/A
1000 mV, 1 kHz
± 8.5 mV
1000 mV, 10 kHz
± 47 mV
1000 mV, 20 kHz
N/A
1000 mV, 30 kHz
± 47mV
1000 mV, 100 kHz
N/A
Error from nominal 1 year
U1252B
± 0.24 mV
± 0.415 mV
± 1.695 mV
N/A
± 1.695 mV
± 2.25 mV
± 2.25 mV
± 4.15 mV
± 16.95 mV
N/A
± 16.95 mV
± 6.5 mV
± 11.5 mV
± 11.5 mV
N/A
± 47.0 mV
6
129