Table 2-3 Audible Continuity Measurement Range - Agilent Technologies U1252B User's Manual And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1252B:
Table of Contents

Advertisement

2 Making Measurements
36
Audible continuity
In the range of 0–500 Ω, the beeper will sound if the
resistance value falls below 10 Ω. For other ranges, the
beeper will sound if the resistance falls below the typical
values indicated in

Table 2-3 Audible continuity measurement range

Measurement range
500.00 Ω
5.0000 kΩ
50.000 kΩ
500.00 kΩ
5.0000 MΩ
50.000 MΩ
500.00 MΩ
Conductance
Set up the multimeter to measure conductance as shown in
Figure 2-
9. Probe the test points and read the display.
The conductance measurement enables the measurement of
very high resistance of up to 100 GΩ.
As the high- resistance readings are susceptible to noise, you
can capture the average readings by using the Dynamic
Recording mode. Refer to the section
Recording"
on page 50 for more information.
Agilent U1251B/U1252B User's and Service Guide
Table 2- 3
below.
Beeper sound threshold
"Dynamic
< 10 Ω
< 100 Ω
< 1 kΩ
< 10 kΩ
< 100 kΩ
< 1 MΩ
< 10 MΩ