Performance Tests And Calibration; Calibration Overview; Closed-Case Electronic Calibration; Agilent Technologies Calibration Services - Agilent Technologies U1252B User's Manual And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1252B:
Table of Contents

Advertisement

6

Performance Tests and Calibration

Calibration Overview

N O T E

Closed-case electronic calibration

Agilent Technologies calibration services

118
This manual contains procedures to verify the instrument's
performance and adjustment (calibration).
The performance test procedures allow you to verify that the
handheld digital multimeter is operating within its published
specifications. The adjustment procedures ensure that the
multimeter remains within its specifications until the next
calibration.
Make sure you have read the
calibrating the instrument.
This instrument features closed- case electronic calibration.
No internal mechanical adjustments are required. The
instrument calculates correction factors based upon the
input reference value you set. The new correction factors are
stored in the nonvolatile memory until the next calibration
adjustment is performed. The nonvolatile EEPROM
calibration memory is retained even when the power is
switched off.
When your instrument is due for calibration, contact your
local Agilent Service Center to enquire about recalibration
services.
Agilent U1251B/U1252B User's and Service Guide
"Test Considerations"
on page 124 before