Percentage scale of 4 mA to 20 mA
N O T E
Agilent U1251B/U1252B User's and Service Guide
Set up the multimeter to measure percentage scale as shown
in
Figure 2-
4. Probe the test points and read the display.
• Press
to select percentage scale display. Ensure that
is shown on the display.
• The percentage scale for 4 mA to 20 mA or 0 mA to 20 mA is calculated
using its corresponding DC mA measurement. The U1251B and U1252B
will automatically optimize the best resolution according to
below.
• Press
to change the measurement range.
The percentage scale for 4 mA to 20 mA or 0 mA to 20 mA
is set to two ranges as follows:
Table 2-2 Percentage scale and measurement range
Percentage scale (4 mA to 20 or 0 mA to 20 mA)
Always auto range
999.99%
9999.9%
Making Measurements 2
or
Table 2-2
DC mA auto or manual
range
50 mA, 500 mA